Dynalyzer – High Voltage Unit
OPERATION MANUAL
11
The setup as shown in Figure 4 may also be employed in grid-pulsed applications to accurately
determine anode kVp and x-ray tube current. This technique does not measure cathode kVp or
filament current because the cathode side of the Dynalyzer is not connected.
Series Parallel Connection
This method can only be used with x-ray generators having auxiliary high-voltage
receptacles capable of simultaneous activation.
This setup shown in Figure 7 enables the measurement of all parameters except filament
current and eliminates the possible adverse effect of an additional test cable in the filament
circuit.
NOTE:
The Dynalyzer IV has been designed for remote reading to minimize x-ray danger to
the operator.
CAUTION:
It is recommended that the High Voltage Unit and the Digital Display or
oscilloscope (if used) be connected to the same grounded outlet or
additional noise may be observed on the mA output.