Measurement Output Selection –
This table
shows on which
screen the
output will be
shown.
* Any of the outputs
can be shown on
screens C1 and C2.
Value
Screen* Definition
Digitizer
Air Pressure
Duration
FWHM Duration
Pulse Count
Multi-Sensor
Dose
MS1
Dose Integral
Dose / mAs
mA
Dose / mAs
Dose / Pulse
Dose / Pulse
Dose Rate
MS2
Dose Rate
Dose Rate Max
Maximum dose rate encountered
Dose Ratio AGMS / IC
Multi-sensor dose divided by ion-chamber dose
Filtration
MS2
Measured beam-filtration thickness.
HVL
MS1
Half value layer result.
kV
MS1
kV average
kVp
MS2
Average peak voltage
Ion Chamber
Dose
IC
Dose Integral
Dose / Pulse
Dose / Pulse
Dose Rate
IC
Dose Rate
Dose Rate Max
Maximum dose rate encountered
Temperature
Temperature measured by ion chamber.
The atmospheric corrections compensate for the effects of variations in the current
temperature (T in C) and pressure (P in kPa) on an unsealed ion chamber and uses the
following equation: Displayed value = (101.3/P) * ((T + 273.15)/295.15) * Instrument value
Exposure duration from start to stop trigger. If pre-pulse detection was activated,
duration is the duration of the main signal without pre-pulse.
Full-Width at Half-Maximum duration. The width of the time interval within which
the signal is at least half the peak signal.
Measurement pulse count.