SGA01 Strain Gauge Amplifier 

 

 
 

30/04/2008

 

Page 5 of 6 

Calibration 

Calibration of the Strain Gauge Amplifier takes place in two steps, or two steps per channel if more than one channel is being used: offset calibration and 
scale calibration. 
 
First calibrate the offset by subjecting the load cell to the conditions that relate to a zero value (typically no load). With the required load applied, adjust the 
“Offset” screw for that channel until the output voltage reads 0V. 
 
Next, calibrate the scale by subjecting the load cell to a known non-zero value. With the required load applied, adjust the “Scale” screw for that channel until 
the output voltage gives the voltage desired for the applied load. 
 
Repeat for the second channel if necessary. 
 

Specification 

Supply Voltage 

6V – 18V DC 

 

 

CMRR

 

90dB typical (77dB minimum) 

 

 

Bridge Supply Voltage 

5V 

±

2mV 

 

 

Output Range 

±

9V 

 

 

Gain 

 

25 – 1000 (adjustable) 

 

 

Operating Temperature 

-40

°

C – +85

°

 

 

 