  Using the buttons to program the temperature controller: 

a.

 

Setting  Mode:  After  open  the  machine,  pressing  set  for  3  seconds  to  enter  the  “setup”  mode.    The  display  will 

show  “ts”  and  the  “set”  light  will  glow.    Press  ▲  or  ▼  in  this  mode  to  select  the  parameter  code:    Parameter 

codes include: TS,TD,HS,LS,AU,AL,AD,AC,OT,OU. 

b.

 

Display or change the parameter:    After choosing the parameter code, press set to display parameter.    Then press 

▲ to increment the value or press ▼ to decrement the value.    You can hold the button depressed for some time to 

increase  or  decrease  automatically.    When  the  parameter  is  as  desired,  then  press  set  button  to  store  the  value.   

After  saving  the  value  to  the  code,  the  display  will  return  to  the  menu  to  show  the  next  code.    To  exit  the 

parameters and exit, choose parameter ”OU” and then press set .    The temperature controller display will leave the 
“Setup” and return to the normal mode. 

c.

 

In the “Setup” mode, if the user does not press any button within 15 seconds, then the temperature controller will 

exit the mode, save the parameter, and return to normal mode.    The display will show the current temperature. 

5.

 

Displaying the highest and lowest temperature: when the temperature arrives at the set temperature, start to make record 

of the highest or lowest temperature ever reached. Please press Log button to see the highest or lowest temperature ever 

reached,  if  keep  pressing  Log  button  for  5  seconds,  the  record  will  become  the  temperature  of  cool  water  and  make 

record again. 

6.

 

Quick Setting:    The operator can amend the set temperature(ts) with few simple steps, as follows: 

1)

 

Press ▼ for 3 seconds to enter setting mode, at this moment the set light will glow. 

2)

 

Press ▲ or ▼ to amend the temperature. 

3)

 

When  the  temperature  is  as  desired,  press  Set  or  without  pressing  any  button  within  5  seconds,  the  system  will 

save the new temperature and come back to the normal condition. 

7.

 

Locking the parameters: 

a.

 

Press Set and ▼ at the same time for two seconds.    If the display shows “LC”, then the parameter is locked. 

b.

 

If you want to unlock the parameter, then Press Set and ▼ at the same time for two seconds.    The display shows 
“UL” to indicate parameter is unlocked. 

8.

 

Resetting all parameters to the factory values: 

a.

 

Pressing Set and ▼ at the same time then turn on the electricity. The temperature controller will start up come back 

with  the  original  parameter  values.    The  display  will  show  “rS”.    After  2  seconds,  the  controller  will  restart 

automatically. 

9.

 

The LED light and warning codes: 

1)

 

“Comp” Is glowing when the compressor moves.    Is off when the compressor is stopped.    Is blinking when the 

compressor is in delay mode. 

2)

 

“Defr” Is glowing when the mode of taking off the frost. 

3)

 

“Alarm” Is glowing when there is a fault or warning. 

4)

 

“Set” Is glowing when in the “Setup” mode. 

5)

 

“bfaults” code( when is bellow saturation, the compressor will run 15 minutes and stop 15 minutes) 

6)

 

“EO” Identifies revise temperature sensor may be damaged.( send back to be repaired) 

7)

 

“E1”  Identifies  the  water  sensor  may  be  damaged  (Check  the  sensor  wires  to  see  they  are  contacted,  make  sure 

water is actually flowing.) 

8)

 

“EE” Identifies the value that was set into the memory parameter was invalid. (Reset the electricity again or change 

the parameter according to the factory original value) 

9)

 

The following warning codes will reset after the temperature reaches the set-point temperature for the first time. 

a)

 

“UA”    The temperature of cool water exceeds the limited temperature. 

b)

 

“LA”    The temperature is below the lowest temperature. 

c)

 

Warning output:(output is on one second/off one second) 

d)

 

When the warning delay time arrives, the warning output  will  turn  on.    The operator can press  ▼  to  mute  the 

warning output temporarily. Pressing this button again will toggle the warning output back on. 

   