background image

Page 3

         VIN1 FS: 3716

         VIN2 FS: 3715

         VIN3 FS: 4094

         VIN4 FS: 4095

          FORMAT: RMY POLLED

       POLL ADDR: a

     OUTPUT RATE: 15 HZ

            BAUD: 9600

OFFSET X,Y,RATIO: -251, 75, 847

 COMPASS DAMPING: SLOW

  WATCHDOG TIMER: ENABLED

1.2.6 POLLED MEASUREMENTS

When the serial output format is RMY POLLED, the 32500 sends
data only when it receives a polling command that contains the
unique address of the unit being polled. The command for polling is:

Ma!

where a = address character.

See the  SERIAL FORMAT DIAGRAM for details on the data sent
from the 32500 when polled.

1.3 MAINTENANCE

Like all quality measuring instruments, the 32500 Electronic Com-
pass should be periodically recalibrated. The recalibration period de-
pends on the application but once per year is typical. The unit may
be sent to the factory or your distributor for recalibration. A copy of
the calibration procedure is available to qualified technicians.

In order to compensate for local magnetic conditions, the compass
calibration can take place only at the site where it is installed. 

If the

installation is changed, the compass must be recalibrated for
the new conditions.

 In addition to changing the actual installation,

local conditions may also change by the introduction of iron masses
or electrical machinery and thus warrant recalibration.

1.4   WARRANTY

This product is warranted to be free of defects in materials and con-
struction for a period of 12 months from date of initial purchase. Li-
ability is limited to repair or replacement of defective item. A copy of
the warranty policy may be obtained from R. M. Young Company.

CE COMPLIANCE

This product has been tested and shown to comply with European
CE requirements for the EMC Directive. Please note that shielded
cable must be used.

Declaration of Conformity

Application of Council Directives:

 89/336/EEC

Standards to which

 

Conformity is Declared:

EN 50081-1
EN 55022 (CISPR 22 class A)
EN 50082-1 (IEC 801-2, 3, 4)

Manufacturer's Name and Address: 

R. M. Young Company

Traverse City, MI, 49686, USA

Importer's Name and Address:

 See Shipper or Invoice

Type of Equipment:

 Meteorological Instruments

Model Number / Year of Manufacture:

 32400/32500/2000

I, the undersigned, hereby declare that the equipment
specified conforms to the above Directives and Standards.

Date / Place:

 Traverse City, Michigan, USA September 1, 2000

David Poinsett
R & D Manager, R. M. Young Company