background image

Page 2

05305L-90(J)

Initial installation is most easily performed using two people; one 

to adjust instrument position and the other to observe an indicating 

device. After initial installation with the included orientation ring, the 

instrument can be removed and returned to its mounting without 

re-alignment. Install the Wind Monitor following these steps:

1. MOUNT  WIND  MONITOR

  a) Place orientation ring on mounting post. Do Not tighten band 

clamp yet. (Orientation ring may be omitted when portable 

tripod is used.)

  b) Place Wind Monitor on mounting post. Do Not tighten band 

clamp yet. 

2. CONNECT SENSOR CABLE.

  a) Connect sensor cable according to wiring diagram located at 

back of manual.

3. ALIGN VANE

  a) Connect instrument to indicator.

  b) Choose a known wind direction reference point on the 

horizon.

  c) Sighting down instrument center line, point nose cone at 

reference point on horizon.

  d) While holding vane in position, slowly turn base until indicator 

shows proper value.

  e) Tighten mounting post band clamp.

  f)  Engage orientation ring indexing pin in notch at instrument 

base.

  g) Tighten orientation ring band clamp.

CALIBRATION

The Wind Monitor is fully calibrated before shipment and should 

require no adjustments. Recalibration may be necessary after 

some maintenance operations. Periodic calibration checks are 

desirable and may be necessary where the instrument is used in 

programs which require auditing of sensor performance.
Accurate wind direction calibration requires a Model 18112 Vane 

Angle Bench Stand. Begin by connecting the instrument to a signal 

conditioning circuit which has some method of indicating the wind 

direction value. This may be a display which shows wind direction 

values in angular degrees or simply a voltmeter monitoring the 

output. Orient the base so the junction box faces due south. 

Visually align the vane with the crossmarkings and observe the 

indicator output. If the vane position and indicator do not agree 

within 5°, it may be necessary to adjust the potentiometer coupling 

inside the main housing. Details for making this adjustment appear 

in the MAINTENANCE, POTENTIOMETER REPLACEMENT, 

outline, step 7.
It  is  important  to  note  that  while  the  sensor  mechanically 

rotates through 360°, full scale wind direction signal from the 

signal conditioning occurs at 355°. The signal conditioning 

electronics  must  be  adjusted  accordingly.  For  example,  in 

a circuit where 4 to 20 mA represents 0° to 360°, the output 

must be adjusted for 19.8 mA when the instrument is at 355°.   

[((355°/360°)  x 16 mA) + 4 mA]. 
Wind speed calibration is determined by propeller pitch and the 

output characteristics of the transducer. Calibration formulas 

showing wind speed vs. propeller rpm and output frequency are 

included below. Standard accuracy is ± 0.2 m/s (0.4 mph). For 

greater accuracy, the sensor must be individually calibrated in 

comparison with a wind speed standard. Contact the factory or 

your supplier to schedule a NIST (National Institute of Standards & 

Technology) traceable wind tunnel calibration in our facility.
To  calibrate  wind  system  electronics,  temporarily  remove 

the  propeller  and  connect  an Anemometer  Drive  (18802 

or equivalent)  to the propeller shaft. Apply the calibration 

formula to the calibrating motor rpm and adjust the electronics 

for the proper value. For example, with the propeller shaft 

turning at 3600 rpm, adjust an indicator to display 18.4 m/s   

[3600 rpm x 0.00512 (m/s)/rpm = 18.4 m/s.]

Details on checking bearing torque, which affects wind speed and 

direction threshold, appear in the following section.

CALIBRATION FORMULAS

Model 05305L Wind Monitor w / 08254 Propeller

WIND SPEED  vs  PROPELLER RPM

m/s

=

0.00512 x rpm

knots

=

0.00995 x rpm

mph

=

0.01145 x rpm

km/h

=

0.01843 x rpm

WIND SPEED  vs  mA OUTPUT

m/s

=

( 6.250 x mA)-25

knots

=

(12.125 x mA)-48.5

mph

=

(13.980 x mA)-56

km/h

=

(22.500 x mA)-90

WIND DIRECTION  vs  mA OUTPUT

DEGREES

=

(22.5 x mA)-90

MAINTENANCE

Given  proper  care,  the  Wind  Monitor  should  provide  years  of 

service. The only components likely to need replacement due to 

normal wear are the precision ball bearings and the wind direction 

potentiometer. Only a qualified instrument technician should perform 

the replacement. If service facilities are not available, return the 

instrument to the company. Refer to the drawings to become familiar 

with part names and locations. The asterisk * which appears in 

the following outlines is a reminder that maximum torque on all 

set screws is 80 oz-in.

POTENTIOMETER    REPLACEMENT

The potentiometer has a life expectancy of fifty million revolutions.  

As it becomes worn, the element may begin to produce noisy signals 

or become non-linear. When signal noise or non-linearity becomes 

unacceptable, replace the potentiometer. Refer to exploded view 

drawing and proceed as follows:
1. REMOVE MAIN HOUSING

  a)  Unscrew nose cone from main housing. Set o-ring aside for 

later use.

  b)  Gently push main housing latch.

  c)  While pushing latch, lift main housing up and remove it from 

vertical shaft bearing rotor.

2. UNSOLDER TRANSDUCER WIRE

  a)  Remove junction box cover, exposing circuit board.

  b)  Remove screws holding circuit board.

  c)  Unsolder  three  potentiometer  wires  (white,  green,  black), 

two wind speed coil wires (red, black), and earth ground wire 

(red) from board.

3. REMOVE POTENTIOMETER

  a)  Loosen set screw on potentiometer coupling and remove it 

from potentiometer adjust thumbwheel.

  b)  Loosen set screw on potentiometer adjust thumbwheel and 

remove it from potentiometer shaft.

  c)  Loosen two set screws at base of transducer assembly and 

remove assembly from vertical shaft.

  d)  Unscrew potentiometer housing from potentiometer 

mounting & coil assembly.

  e)  Push potentiometer out of potentiometer mounting & 

coil assembly by applying firm but gentle pressure on 

potentiometer shaft. Make sure that the shaft o-ring comes 

out with the potentiometer.  If not, then gently push it out 

from the top of the coil assembly.

Содержание 05305L

Страница 1: ...MENTS INSTRUCTIONS R M YOUNG COMPANY 2801 AERO PARK DRIVE TRAVERSE CITY MICHIGAN 49686 USA TEL 231 946 3980 FAX 231 946 4772 WEB www youngusa com P N 05305L 90 REV J111215 WIND MONITOR AQ WITH 4 20mA...

Страница 2: ...ns As a general rule the air flow around a structure is disturbed to twice the height of the structure upwind six times the height downwind and up to twice the height of the structure above ground For...

Страница 3: ...ated in comparison with a wind speed standard Contact the factory or your supplier to schedule a NIST National Institute of Standards Technology traceable wind tunnel calibration in our facility To ca...

Страница 4: ...n through front of main housing and turn potentiometer adjust thumbwheel until signal conditioning system indicates proper value d Tighten set screw on potentiometer coupling 8 REPLACE NOSE CONE a Scr...

Страница 5: ...se This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme Ia norme NMB 001 du Canada EN55011 CISPR 11 Group 1 Class B device Class B equipment is suitable for use in domestic es...

Страница 6: ...TION FORMULAS Model 05305L WIND SPEED vs mA OUTPUT m s 6 250 x mA 25 knots 12 125 x mA 48 6 mph 13 980 x mA 56 km h 22 500 x mA 90 WIND DIRECTION vs mA OUTPUT DEGREES 22 5 x mA 90 CABLE WIRING DIAGRAM...

Страница 7: ...Page 6 05305L 90 J BEARING REPLACEMENT POTENTIOMETER ADJUSTMENT...

Страница 8: ...G 05330D TRANSDUCER ASSEMBLY 05324 2 VERTICAL SHAFT BEARING 05187 PROPELLER NUT 1 4 20 05164C PROPELLER SHAFT W HUB 6 32 x 1 8 SET SCREW 5 05127 MAIN HOUSING LATCH 6 32 x 3 8 FLAT HD SCREW 05126C VERT...

Страница 9: ......

Страница 10: ...Accessories Model 18802 Anemometer Drive Model18112 Vane Angle BenchStand Model 18310 Propeller Torque Disc Model18331Vane Torque Gauge Model 18301 Vane Alignment Rod Model 18212 Vane Angle Fixture T...

Страница 11: ...ne Torque Gauge checks vane bearing torque of the Wind Monitor family sensors Slip the fixture over the main housing and make simple yet accurate vane torque measurements Charts relating vane torqueto...

Отзывы: