Problem

Causes and Solutions

No display appears on the 
display screen.

No power.

Measurement values 
appear too high or too low.

Check the battery installation for proper 
placement of the batteries polarities.

Replace all the worn batteries with 4 new 
ones.

Blood pressure varies constantly. Many 
factors including stress, time of day, how 
you wrap the cuff, may affect your blood 
pressure. Review the sections "Proper Use 
of the Arm Cuff" and "Take a Measurement".

Eliminate Fault

Symbol

Cause

Correction

When the batteries 
power are too low.

The course of 
inflating appears 
error.

When measurement 
fails.

Replace all of the worn batteries 
with new ones.

Wrap the cuff correctly and tightly.

Ensure the connection is close
between the air plug and the unit.

Inflate over again after ensuring.

Measure over again according to 
correct way.

Do not move your arm and body,
and keep quiet.

Error Indicators

The following symbol will appear on the display when measuring abnormally.

Exceptional Situations

Exceptional Situations

*Note!

When a person’s systolic and diastolic pressures fall into different categories, 
the higher category should apply.

After each measurement is completed, the LCD display will show your 
position automatically on the six segments of the bar indicator which 
corresponds to World Health Organization (WHO) Blood Pressure Indicator.

Reference Material: Journal of Hypertension 1999, Vol 17 No.2

Reference Material: Journal of Hypertension 1999, Vol 17 No.2

80

85

90

100

110

Diastolic Blood Pressur (mmHg)

120

130

140

160

180

Systolic Blood Pressure (mmHg)

Optimal

blood pressure

(target value)

Normal blood pressure

Normal systolic value

Mild hypertension

Moderate hypertension

Severe hypertension

Содержание BP101U

Страница 1: ...tal Blood Pressure Monitor Please read this instruction manual before using your unit Please keep this instruction manual safe for future use Instruction Manual Lead a Healthy Life 0197 QUOTICARE QUOT...

Страница 2: ...ant Safety Notes 3 Warning and safety notices 4 Konw Your Unit 5 6 Quick Reference Guide 7 Battery lnstallation Replacement 8 Adjusting Settings 1 To set year 9 2 To set date and time 10 3 To set time...

Страница 3: ...ure Self diagnosis and treatment using measured results may be dangerous Follow the instructions of your healthcare provider Operate the unit only as intended Don t use for any other purpose The unit...

Страница 4: ...cation of the CUFF and its pressurization on the arm on the side of a mastectomy Regarding the need to check for example by observation of the limb concerned that operation of the AUTOMATED SPHYGMOMAN...

Страница 5: ...pper arm perimeter If your arm cuff is broken or not functional please use a new cuff A new arm cuff does not include an air plug Please continue to use the old air plug on the new arm cuff Hose Air P...

Страница 6: ...eries should be carried out in accordance with the national regulations for the disposal of batteries Press the MEM button to increase one year Long press the button and the value quickly changes Adju...

Страница 7: ...justing Settings 4 To set unit 4 To set unit 11 Adjusting Settings Adjusting Settings Then press the MEM button again to change the current the 12 hour or the 24 hour Press the SET button again and Co...

Страница 8: ...eart level throughout the measurement 6 Keep a record of your blood pressure and pulse readings for your physician A single measurement does not provide an accurate indication of your true blood press...

Страница 9: ...rm The bottom of the cuff should be approximately 1 2cm above your elbow 6 5 Put your arm through the loop then pull it up to the position of your upper arm Proper Use of the Arm Cuff Proper Use of th...

Страница 10: ...t symbol flashes once a pulse is detected Take a Measurement Take a Measurement The cuff is deflated automatically and all of the measurement results are stored in the memory with date and time The sy...

Страница 11: ...alue Press the MEM Memory button to view the latest measurement value with the measuring date and time Press the SET Setting button to view the oldest recorded value with the measuring date and time L...

Страница 12: ...er pressure or lower value represents the blood pressure produced by relaxation of the heart muscle Blood Circulation IHB Introduction IHB Introduction When the unit detects an irregular rhythm two or...

Страница 13: ...t Inflate over again after ensuring Measure over again according to correct way Do not move your arm and body and keep quiet Error Indicators The following symbol will appear on the display when measu...

Страница 14: ...kind yourself If a defect occurs Please contact the local distributor The unit should be cleaned with a soft dry cloth Use a soft moistened cloth and soap to clean the arm cuff Substitution of a cuff...

Страница 15: ...etic environment guidance Electromagnetic environment guidance The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below The customer or the user of Dig...

Страница 16: ...nd mobile RF communications equipment should be used no closer to any part of the Digital Blood Pressure Monitor including cables than the recommended separation distance calculated from the equation...

Страница 17: ...abor only under normal operations A transportation fee or freight fee that may be incurred will be the owner s responsibility Any defect resulting from natural causes e g flood hurricane etc is not wi...

Отзывы: