Q150T Sample Preparation System
10473 - Issue 5
48
Q150T - Instruction Manual
4.
Edit other parameters as required.
5.
Back on the
Maintenance
page, click on the
OK
button to confirm the changes or
on the
Cancel
button to restore the previous values.
6.2.5
Hardware
This option allows you to change the settings defining the configuration of your
instrument. The Q150T uses these settings in all profiles of type:
Timed Sputter
and
FTM Terminated Sputter
.
To change the settings, tap on
Hardware
and then tap on the
Edit
button. The
Hardware
properties are displayed (see
Figure 6-7
and
Table 16
).
Figure 6-7.
Hardware settings (single target)
Table 16
Hardware settings
Name
Default
Value
Minimum
Value
Maximum
Value
Comment
Target A
Target material, if installed.
Target B
Target material, if installed.
Target C
Target material, if installed.
Vacuum Gauge
Type
Read only: displays gauge type
Use Turbo
Pumping*
Yes
No
Yes
If
No
the turbo pump will not be
turned on during the pumping of
the chamber
FTM Present
No
No
Yes
Turbo Present
Yes
No
Yes
Stage Rotate
Yes
No
Yes
Set the state of the stage unless
overridden in the process profile
Stage Rotate
Speed
Sets the speed of the stage
unless overridden in the process
profile
Available Targets*
1
1
3
Sets the number of auxiliary
sputter inserts allowing the whole
insert to be changed making the
use of multiple materials easier
* Admin group level required to edit this property
1.
To edit a parameter, tap on its value.
2.
Change the value as required. The instrument displays a dialog box for editing
purposes, with either a numeric keypad or keyboard as required. The dialog box
also shows the allowed values for the parameter.