0.001-6 GHz Dual Channel Upconverter
IEEE 488.2 Command Reference
6.7 *RCL
<
NRf
>
The
*RCL
(ReCaLl) command restores a previously saved state from non-volatile memory. In addition to
factory default state 0, there are 5 re-writeable memory locations, specified by choosing an index between
1 and 5. The restored parameters are the
Tune Frequency, Channel 1 Attenuator 1 Setting, Channel 1 At-
tenuator 2 Setting, Channel 1 Attenuator 3 Setting, Channel 2 Attenuator 1 Setting, Channel 2 Attenuator
2 Setting, and Channel 2 Attenuator 3 Setting
. For a detailed description of the parameters saved, see
the relevent command descriptions in this manual. When the *RCL command is called, these parameters
will be applied to the device.
Syntax
*
RCL
Space
NRf
Allowed Values
The
NRf
parameter can be any integer in the range of 0 to 5.
Error Message
If the parameter is not in the recognized format, error
-102, "Syntax error"
occurs.
If the parameter is less than 0 or greater than 5, error
-222, "Data out of range"
occurs.
Quonset Microwave
Revision 1.2.0
68