Lead Check

1. Select the Menu by pressing the 

Up/Right button continuously 

for several seconds. “Lead Check” will appear.

2. From the Menu, select 

Lead Check by pressing the Enter key.

3. A graph depicting the impedance measured at the Right Arm 

(RA), Left Arm (LA), Left Leg (LL) and optional V leads is displayed 

from left to right in vertical columns on the screen. The higher the 

bar, the better the contact is between the skin and the electrode.

4. A full-bar graph means optimal high quality and good electrode 

contact. For good quality transmissions, the bars should reach or 

exceed the horizontal line on the display. A low-bar graph means 

poor quality and high electrode impedance. The skin preparation 

should be checked for improvement and, if necessary, the 

electrode(s) should be replaced.

5. Once acceptable impedance levels are ver ed, press any of the 

three keys to return to the Lead Check menu.

6. Press the 

Down key to scroll to Done. Press the Enter key.

Lead Failure

A lead failure is displayed as a lead designator on the LCD display. If 

the patient cable is not attached, the LCD will display all leads as 

disconnected.

Battery Voltage Indicator

The X12+ is powered with a single AA alkaline battery that requires a 

minimum of 1.0 volts to operate. When the battery contains su cient 

voltage, the main menu displays an image of a battery in the upper 

right corner showing the current battery voltage in increments of 

100%, 75%, 50%, and 25%. If a battery with unknown voltage is 

inserted and the LCD menu does not appear, insert a new battery.

When the battery indicator shows a voltage of 25%, discard the 

battery and insert a new battery into the battery compartment.

REF: 70-00256-01 Rev C1

Attaching Patient Cable

The Patient Cable consists of a connector block, main cable and four 

ve leadwires connected to the main cable. Each leadwire 

terminates in a snap connector. The leadwires are positioned on the 

main cable to follow the contour of the torso.

Insert the connector block into the input connector on the side of the 

transmitter, making sure that it is parallel to the transmitter input 

connector.

Reorder Information (www.mortara.com)

Part No.

Description 

000310-001

Quik-Trace multipurpose electrodes (30/pouch)

036817-013

Transmitter lead set, 4-wire, AHA 

036817-014

Transmitter lead set, 4-wire, IEC

036817-003

Transmitter lead set, 5-wire, AHA

036817-004

Transmitter lead set, 5-wire, IEC

042174-001 

Pouch and belt for X12+ Transmitter