MedTrack CR60 Rehabilitation Treadmill Service Manual
Repair/Replacement and Calibration 4-29
7. Rotate the grade pot shaft with a screwdriver until the grade display
reads
0.0
.
8. Tighten the setscrew and verify that the display still reads
0.0
.
9. Press
Clear
to exit open-loop mode.
10. Verify that the treadmill operates through its full grade range (0 –
15%) and that the treadmill is nearly all the way to the top of the rack
gear when the display reads 15%.
11. Turn off the power and disconnect the power cord from the power
outlet.
12. If no other service is required, replace the treadmill hood as described
in step 10, page 4-5.
Speed Calibration
The speed of the belt can be checked for accuracy. The display and
control are digitally controlled and cannot be adjusted. See “Test E. Test
Speed Operation” on page D-3.
Grade Calibration
The treadmill grade is controlled through the grade pot and grade pot
installation procedure and cannot be adjusted. See “Test D: Test the Grade
Operation” on page D-3.
Содержание MedTrack CR60
Страница 1: ...MedTrack CR60 Rehabilitation Treadmill Service Manual P N 042101 001 Rev A...
Страница 4: ......
Страница 27: ...MedTrack CR60 Rehabilitation Treadmill Service Manual Troubleshooting 3 13 Grade Problems Flow Chart...
Страница 28: ......
Страница 58: ......
Страница 62: ......
Страница 70: ......
Страница 78: ...C 4 Part Numbers MedTrack CR60 Rehabilitation Treadmill Service Manual 13 18 27 27 26 26 26 15 15...
Страница 80: ......
Страница 88: ......
Страница 89: ...E 3 Assembly Drawings Cardiac Stress Treadmills Service Manual Wiring Diagram Low Voltage Treadmill...
Страница 90: ...E 4 Assembly Drawings Cardiac Stress Treadmills Service Manual Wiring Diagram High Voltage Treadmill...