Read Operating Instructions Thoroughly Prior to Operation
Safety Precautions
Read Opera�ng Instruc�ons thoroughly prior to opera�on and observe the following safety precau�ons:
• Use only a grounded outlet that is rated for your model's electrical requirement.
• Do not modify the oven or factory control se�ngs to operate the oven above the stated maximum opera�ng temperature.
•
Exterior surfaces on the 180 models may become hot to the touch when opera�ng at higher set temperatures.
• Conduct periodic maintenance as required.
Do not place vola�le or combus�ble materials into incubators.
This unit is not intended for use with any flammable liquids or vapors, or with chemicals that produce toxic gases.
Safety Precautions
Place the Incubator on a flat surface. Maintain a minimum of 3” (7.6
cm) of airspace around the unit and a minimum of 16” (40.6 cm)
above the unit to allow for the placement of the thermometer.
Install the desired shelves in the unit (see Shelf Installa�on).
Plug the unit into a grounded outlet that is rated for your model's
electrical requirement
.
20
40
60
80
10
0
20
40
60
80
20
0
20
40
60
80
30
0
20
40
60
80
40
0
20
40
60
80
50
0
0
10
20
30
40
50
60
70
80
90
10
0
10
20
30
40
50
60
70
80
90
20
0
10
20
30
Minumum
Required Space
above the
incubator for
thermocouple
placement.
16” (inches)
INCUBATOR SHELF INSTALLATION
Your incubator allows shelves to be adjusted at different heights in the
unit and are easy to install and adjust using the wire shelf brackets. To
install and/or adjust shelves follow the steps below:
THERMOMETER INSTALLATION
If you encounter any shipping damage to the unit and/or are missing any of the units contents, please refer to the
enclosed
DAMAGE/MISSING ITEM REPORT
or call 1800-482-4328.
FIG. 2
FIG. 3
FIG. 1
Remove the glass thermometer from its container and insert
into the rubber grommet using a slow twis�ng mo�on.
(FIG. 4)
I
MPORTANT:
Do NOT modify the provided plug. Do NOT use an extension cord
Use an individual branch circuit for your incubator. For informa�on
on your unit’s electrical specifica�ons and rated voltage, refer to
label on the back of the unit.
FIG. 4
FIG. 5
NOTE:
When inster�ng the
rubber grommet on top of the
unit, ensure that the
narrow
part
of the grommet is facing down.
Grommet will only go in about
1/4 inch into the top port hole.
Align bracket ends with holes located inside the ovens walls
(FIG. 2)
Insert shelf bracket ends into the hole and press down on the bracket
as shown in
(FIG. 3)
. Repeat proccess on the opposite side of the wall.
Place shelf on top the the brackets.
PAGE 2
WARNING:
CAUTION:
INCUBATOR SET-UP
AVOID FORCING GLASS THERMOMETER IN GROMMET, THIS CAN CAUSE
THERMOMETER TO SHATTER , RISKING A POSSIBILITY OF INJURY.
Insert the rubber grommet/thermometer assembly into the
port through the top of the incubator. The thermometer
should be insterted with the narrow part of the grommet
facing down. Insert by using a slow twis�ng mo�on and allow
the thermometer to extend into the incubator.
(FIG.5)
CAUTION:
Mantain a minimum of 3 inch
spacing around the unit.
UNPACKING AND CONTENTS
Unpack your incubator and inspect it for any shipping
damage. Verify that the (4) rubber feet at the bo�om of the
unit are not missing. Make sure you have received all the
items shown on this sec�on.
The design of the parts and
accessories may differ from the pictures shown. Their design is
subject to change without notice for product improvement.
Fixed Shelf (X1)
Adjustable Shelf (X1)
Shelf Bracket
(X2)
SpiritThermometer
(X1)
Thermometer
Grommet (X1)