Page 36 of 57
QC Results
The displayed results are stored and can be recalled with the Review QC Results function in the
Review Data menu.
Select OK to complete the QC process and return to the Main Menu.
If the QC controls do not pass, repeat the external control testing process or contact the
Supervisor or Quidel Technical Support at 800.874.1517 (in the U.S.) for assistance before testing
patient specimens. See page 4 for additional contact information.
Run Patient Specimens
In the Main Menu, select RUN TEST to get back to the Start Test screen.
Before entering any User ID or specimen information:
Select the desired Development mode: WALK AWAY or READ NOW (See description of the
WALK AWAY and READ NOW development modes on page 13).
The current (or default) development mode will be shown at the top of the Start Test
screen (Figure 13). If the default mode is the desired mode, no changes are required.
Figure 13
If the Supervisor has set the default mode to Locked WALK AWAY mode, then User will
only be allowed to test in the WALK AWAY mode.
IMPORTANT: Refer to the assay-specific Package Insert for each specific test for instructions on
preparing Test Cassettes with patient samples.
Содержание Sofia