Quidel Sofia Influenza A+B Скачать руководство пользователя страница 6

 

Reader Display 

Interpretation 

Flu A:   
Flu B:   
Procedural Control:   

Positive Test for Flu A 
(influenza A antigen present) 

Flu A:   
Flu B:   
Procedural Control:   

Positive Test for Flu B 
(influenza B antigen present) 

Flu A:   
Flu B:   
Procedural Control:   

Positive Test for both Flu A and Flu B 
(influenza A and B antigen present) 

Flu A:   
Flu B:   
Procedural Control:   

Negative Test for Flu A and Flu B  
(no antigen detected) 

Flu A:  
Flu B:  
Procedural Control:   

Result Invalid 

 

From the main menu, 
select Run QC. 

Follow the prompts on the screen. Scan the QC Card (located on the kit box). 

Sofia 2 prompts the user to select the desired mode (WALK AWAY or READ NOW). 
Run the External Control Swabs. 

Follow the Swab Test procedure of this Quick Reference Instructions. First test the 
Positive Swab, followed by the Negative Swab. 

After both the Positive and Negative Swabs have been run, the results will be displayed as 

 or 

 

 

INTENDED USE 

The Sofia Influenza A+B FIA employs immunofluorescence to detect influenza A and influenza B viral nucleoprotein antigens in direct nasal swab, nasopharyngeal swab, and nasopharyngeal 
aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimen in transport media from symptomatic patients. This qualitative test is intended for use 
as an aid in the rapid differential diagnosis of acute influenza A and influenza B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive and it 
is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not 
be used as the sole basis for treatment or other patient management decisions. The Sofia Influenza A+B FIA may be used with Sofia or Sofia 2. 

 
ASSISTANCE 

If Sofia, Sofia 2 or the assay do not perform as expected, contact Quidel Technical Support 800.874.1517 (in the U.S.), 858.552.1100 (outside the U.S.), [email protected], or 
your local distributor. 

 

Sofia 2 Interpretation of Results 

Sofia 2 External Quality Control 

(External Positive and Negative Swabs are supplied in the kit)

 

Quidel Corporation 

San Diego, CA 92121 USA 

quidel.com 

 

1219205EN00 (06/17) 

 

When the test is complete, the results will be displayed on the Sofia 2 screen. Test Lines will not 
be visible to the naked eye. 

Results:

 The Sofia 2 screen will display results for the procedural control as being 

 or 

. If the 

control is 

, retest with a new patient sample and new Test Cassette. 

 

Reference the Package Insert for 

Warnings and Precautions,

 

Specimen Collection and Handling,

 and 

Quality Control

Study the Package Insert and User Manual thoroughly 
before using Quick Reference Instructions. This is not a 
complete Package Insert. 

 

Содержание Sofia Influenza A+B

Страница 1: ... bulb b Still squeezing place the Pipette tip into the sample c With the Pipette tip still in the sample release pressure on bulb to fill the Pipette Firmly squeeze the top bulb to empty the contents of the Small Clear 120 µL Fixed Volume Pipette into the Test Cassette sample well Extra liquid in the overflow bulb is OK NOTE The Fixed Volume Pipette is designed to collect and dispense the correct ...

Страница 2: ...he collection cup Firmly squeeze the top bulb to empty the contents of the Large Pink 250 µL Fixed Volume Pipette into the Reagent Tube Extra liquid in the overflow bulb is OK NOTE Once the sample is added to the Reagent Tube vigorously mix prior to adding the sample to the Test Cassette NOTE The Fixed Volume Pipette is designed to collect and dispense the correct amount of patient sample Discard ...

Страница 3: ...mended development time of 15 minutes RUN TEST 1 Input the User ID with the barcode scanner or enter the data using the key pad NOTE If you scan the wrong barcode use the Arrow Keys on Sofia to re highlight the field and simply rescan using the correct barcode The previous one will be overwritten with the right barcode 3 Press Start Test and the Sofia drawer will open 4 Verify that the correct mod...

Страница 4: ...Flu B Positive Procedural Control Valid Positive Test for both Flu A and Flu B influenza A and B antigen present Flu A Negative Flu B Negative Procedural Control Valid Negative Test for Flu A and Flu B no antigen detected Flu A Invalid Flu B Invalid Procedural Control Invalid Result Invalid From the main menu select Run QC Follow the prompts on the screen Scan the QC Card located on the kit box So...

Страница 5: ...in stable for an additional 15 minutes after the recommended development time of 15 minutes RUN TEST 1 Input the User ID with the integrated barcode scanner or manually enter the data using the key pad NOTE If you scan the wrong barcode re highlight the field using the touchscreen on Sofia 2 The previous one will be overwritten with the right barcode 3 Verify that the correct mode WALK AWAY or REA...

Страница 6: ...in the rapid differential diagnosis of acute influenza A and influenza B viral infections The test is not intended to detect influenza C antigens A negative test is presumptive and it is recommended these results be confirmed by viral culture or an FDA cleared influenza A and B molecular assay Negative results do not preclude influenza virus infections and should not be used as the sole basis for ...

Отзывы: