SEATING SECTION
The Recline Actuator will not run, or runs erratically
Complete the following procedures to test the Recline
Actuator. Use the HHP to re-home the chair prior to
completing any tests.
To test the current drain and operation of the en-
1.
coder switches of the Recline actuator, complete
the Fault Isolation Procedures Using Quickie iQ or
HHP located in this manual. Replace the Recline
actuator if the procedure fails. To continue testing,
complete the following procedures.
Locate the QMAC mounted on the back of the
2.
wheelchair (Fig 3.101).
Disconnect the 6-pin female connector (A) leading
3.
to the Recline port of the QMAC (
fi
g 3.102).
Check the armature resistance through the pins
4.
indicated in the photo (Fig 3.103). Nominal resis-
tance is 2 to 4 ohms.
Fig 3.103
Fig 3.102
Fig 3.101
A
The Recline Actuator will not run, or runs erratically
Содержание Rhythm
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical...