
4. To save, press
Save As
key and name the new setting.
5. To send to the instrument, ensure the VI-410 is connected via a USB cable to your pc. Press the
“Send to Instrument” key.
Viewing last calibration
Calibration History
In
calibration history screen, you
can view up to 10 last calibration records.
To access the calibration history screen, follow below:
Menu path: Function>Calibration> >Channel x >Calibration history
Technical Tip
Page 4 04/2009
Figure 1-7: Calibration history screens
1. Select
the
Calibration history
screen.
To access the main menu, press
Altf
key while pressing
Menu/Enter
key.
Select
Function
menu by pressing and press
Menu/Enter
key.
Press and select
Calibration
. Press
Menu/Enter
key.
Press to select the appropriate channel and press
Menu/Enter
key.
Press to select
By Calibration History
and press
Menu/Enter
key.
2. The opened Calibration history screen contains a list of last calibration records. Each calibration is
described by the method of how it was performed either by sensitivity or by measurement. It includes
the date and time of the performed calibration measurement, the obtained calibration factor, and the
calibration result.
Figure 1-8: Calibration history with previous 2 cal. records
NOTE: if calibration measurements were not performed the calibration history screen will state “calibration
history is empty”