Quattroflow EZ-Set Скачать руководство пользователя страница 11

 

 

  

 

                                                                                                                     

 

EZ-Set, Appendix to pump manual of QF1200SU; edition 11/2018; page 11 of 11   

 

4.  Retrofit a pump to EZ-Set  

An existing standard QF1200 single-use pump can be retrofitted quickly and easily to the EZ-Set system. 
For this purpose, the WLC (shaft-bearing-cap) unit must be replaced and the studs screwed in. 

4.1. 

Replacing the shaft-bearing-cap-unit 

 
The  replacement  of  the  shaft-bearing-cap  unit  (item  7)  is  described  in  the  chapter  4.2  of  the  pump 
manual. 

4.2. 

Fasten the studs 

 

 
Fasten the studs (item 6) in the bearing-shell of the ring-drive-unit and tighten them with the help of an 
open-end spanner (A/F 5mm). 
 

 

Quattroflow is a brand of ALMATEC Maschinenbau GmbH.  

ALMATEC is constantly working on improvements of the pump. Modifications of the design or 

materials might be done without prior notice. 

 
 
 

 

Subject to change without notice, 2018-11 (replaces 2018-05)

 

 

ALMATEC Maschinenbau GmbH   

Hochstraße 150-152· D-47228 Duisburg, Germany 

T49 (0) 2065 89205-0 · Fax +49 (0) 2065 89205-40 

www.quattroflow.com · [email protected] 

Содержание EZ-Set

Страница 1: ...ctions QF1200SU EZ Set Please read this appendix and the pump manual carefully before start up of the pump and take care of an appropriate instruction of the operating staff Attention Please regard th...

Страница 2: ...2 3 Dismounting the subassembly 5 2 4 Dismounting the pump chamber 5 3 Assembly 6 3 1 Unscrewing the locking screw 6 3 2 Alignment of the shaft bearing cap unit 6 3 3 Mounting the pump chamber 7 3 4 F...

Страница 3: ...EZ Set Appendix to pump manual of QF1200SU edition 11 2018 page 3 of 11 1 General remarks 1 1 Overview of the components...

Страница 4: ...crew the hand wheel item 4 in rotational direction counter clock wisely 2 3 cycles are enough It is not necessary to screw off the hand wheel completely 2 2 Loosen the bayonet plate Loosen the bayonet...

Страница 5: ...pump manual of QF1200SU edition 11 2018 page 5 of 11 2 3 Dismounting the subassembly Pull off the whole subassembly item 8 to the front side 2 4 Dismounting the pump chamber Pull the pump chamber off...

Страница 6: ...k item 1 and thus the steps 3 1 3 2 3 4 are not absolutely necessary but facilitate the assembly of the pump chamber to fix the position of the cap of the shaft bearing cap unit 3 1 Unscrewing the loc...

Страница 7: ...EZ Set Appendix to pump manual of QF1200SU edition 11 2018 page 7 of 11 Align the cap of the shaft bearing cap unit by the help of the adjusting stick item 1...

Страница 8: ...018 page 8 of 11 3 3 Assembling the pump chamber Assemble the pump chamber by the help of the studs item 6 After that pull out the adjusting stick item 1 3 4 Mounting the locking screw Tighten the loc...

Страница 9: ...on In addition to this the mark on the bajonet plate item 2 should be visible 3 6 Fasten the bayonet plate Fasten the bayonet plate item 2 by turning it clockwisely The bayonet closure snaps in after...

Страница 10: ...10 of 11 3 7 Fasten the hand wheel Fasten the hand wheel item 4 by turning it clockwisely figure 7 Turn the hand wheel until a noticeable drag appears After that the mark of the hand wheel must be tur...

Страница 11: ...cribed in the chapter 4 2 of the pump manual 4 2 Fasten the studs Fasten the studs item 6 in the bearing shell of the ring drive unit and tighten them with the help of an open end spanner A F 5mm Quat...

Отзывы: