V I I . O P E R A T I O N
NOTE: When the batteries approach a discharged state, the first red light will begin to slowly
flash, reminding you the batteries need to be charged immediately!
Off-board Charger/Programming Socket
The off-board charger/programming socket is located on the front of the Remote Plus. If you use an off-board charger, the
charger current should not exceed 12 amps. Contact your Quantum Rehab Specialist for more information.
WARNING! Only chargers with Neutrik NC3MX plugs should be connected to the off-board charger/
programming socket. See your Quantum Rehab Specialist for more information.
NOTE: Changes to the programming can only be made by a Quantum Rehab Specialist.
Controller Communications Cable
The controller communications cable provides the Remote Plus with a connection to the power module.
Sleep Mode
Your Remote Plus controller has a sleep mode feature. Sleep mode is a built-in circuit that automatically shuts off the main
power if the joystick is not moved in any direction for approximately five minutes. The battery condition meter lights on the
keypad indicate sleep mode by blinking once every five seconds. To restore power and continue, push the on/off key twice.
Thermal Rollback
The Remote Plus is equipped with a thermal rollback circuit. This circuit monitors the temperature of the motors, power
module, and remote. In the event that any of them become excessively hot (above 122° F), motor voltage is reduced. For
every degree above 122° F, the voltage is reduced by 5 volts. This reduces your power chairs speed and allows the electrical
components to cool down. When the temperature returns to a safe level, your power chair resumes its normal speed.
Remote Plus Error Codes
In addition to indicating the current state of battery charge, the battery condition meter can also indicate possible problems
with your power chairs electrical system. If any of the battery condition meter lights are flashing rapidly, the controller may
be indicating an error. Error codes are displayed as a number of flashing lights. For instance, if the first light is flashing
rapidly, the battery voltage is nearly depleted. The table below identifies the individual error codes, probable causes, and
possible solutions. If you get one of these error codes, contact your Quantum Rehab Specialist.
FLASHING LIGHTS
DIAGNOSIS
SOLUTION
10
High Battery Voltage
Check batteries.
9
Solenoid Brake Fault
Check motor/brake wiring.
8
Possible Controller Fault
See Quantum Rehab Specialist.
7
Possible Joystick Fault
See Quantum Rehab Specialist.
6
Inhibit Active
Unplug charger. Check connections.
5
Right Motor Wiring Fault
Check right motor wiring.
4
Right Motor Disconnected
Check right motor wiring.
3
Left Motor Wiring Fault
Check left motor wiring.
2
Left Motor Disconnected
Check left motor wiring.
1
Low Battery Voltage
Check batteries/battery wiring.
Содержание 1650
Страница 1: ...Owner s Manual 21 Healey Road Dandenong 3175 Victoria Australia www quantumrehab com ACN 088 609 661 ...
Страница 40: ...40 www pridemobility com Quantum1650 RevA Jan04 N O T E S ...
Страница 41: ...Quantum1650 RevA Jan04 www pridemobility com 41 N O T E S ...
Страница 42: ...42 www pridemobility com Quantum1650 RevA Jan04 N O T E S ...