Part Replacement
Replacing the Height Receiver Board
CubiScan 125 Service Manual
3 0
board correctly, with the power connector in the
bottom-left corner.
9. Screw the replacement transmitter board into place
using the screws from step 7.
10. Reattach the power cable and transmit ribbon, as
. Make sure the
transmit ribbon has been firmly locked into place,
with both tabs securely snapped.
11. Insert the board assembly back into the gate, making
sure that no wires are pinched or disconnected.
Secure the board assembly using the screws from
step 4.
12. Replace the filter and gate cover. When you place the
filter make sure it is clean and free of fingerprints.
Orient the gate cover with the longer side at the top
so that no wires or gaps are visible.
13. Secure the gate cover with the screws from step 2.
14. Power the CubiScan 125 on.
Replacing the Height Receiver
Board
This section describes how to replace the height receiver
board on the CubiScan 125.
Items Needed
•
5/64'' Allen wrench
•
Phillips screwdriver
•
Replacement height receiver board
Complete the following steps to replace the height
receiver board. The height receiver board is located on
the front side of the gate, where the gate handle is
located.
1. Power the CubiScan 125 off.
2. Move the gate so you can access the gate cover.
Loosen the four screws holding the gate cover in
Figure 24
Height Transmitter Board
Figure 25
Height Receiver Gate Cover
Ribbon
cable
Power
cable