Connect the internal laser output to 
a modulator input using the short 
PM patch cord provided with the 
instrument and turn the laser on. Once 
the laser has stabilized, click FORCE 
CALIBRATE in CohesionUI to perform 
the DC bias scan and start the 
Automatic Bias Control (ABC).

Once the ABC has locked, connect the 
RF signal generator to IQTX and turn 
the RF driving signal on. The peak-to-
peak voltage amplitude of the driving 
signal should be set to below 1.8 x 
Vpi for QPSK signals, and around 1.2 
x Vpi for multi-level formats such as 
16 QAM. Typically, the driving signal 
amplitude should be around 200 mV 
to 300 mV. Do not exceed 500 mV as 
this will damage the instrument. 

For detailed operating information, 
refer to the IQTX User Manual 
included on the USB media drive.