15

Quansys Biosciences

Chapter 3: Operation

BEFORE YOU BEGIN

The following procedure is recommended before 

every use.

1.  Wash the glass and glass cover within the 

tray and ensure it is clean and free of debris.  

2.  Check within Q-View Software that the Q-View 

Imager Pro is connected to the PC. 

CLEANING THE TRAY 

SURFACE

If any particulate, liquid, or dried salt residues or 

smudges are left on the glass or the tray, it will 

adversely affect the quality of the image and the 

overall quality of your data. This can be cleaned with 

70 percent ethanol using a lint-free cloth. Do not use 

scouring cleaners or abrasive pads as this can scratch 

the glass.

 

LOADING THE PLATE

When the user pushes the front button, the tray will 

extend. Do not push or apply pressure on the tray 

while it is in the opening or resting position. This will 

adversely affect the performance of the tray as well 

as the alignment with the lens and camera.

     

Содержание Q-View Imager Pro

Страница 1: ...For Research Use Only Version 1 2 Operator s Manual...

Страница 2: ...al in its entirety and train on the device frequently to ensure proper operation If the device is used outside of the instructions herein or modified disassembled or altered the safety of the device c...

Страница 3: ...O NON QUANSYS BIOSCIENCES BRANDED SOFTWARE OR PRODUCTS IS PROVIDED BY THE ORIGINAL MANUFACTURER QUANSYS BIOSCIENCES ASSUMES RESPONSIBILITY FOR THE DELIVERY AND INSTALLATION OF THE Q VIEW IMAGER PRO TO...

Страница 4: ...y Information 3 Safety Precautions 3 Automated Tray 3 Electricity 3 Person Protective Equipment 3 Placement in the Laboratory 3 Temperature and Humidity 3 Ventilation and Dust 4 Space 4 Vibration 4 Q...

Страница 5: ...ernal Controls 11 Front 11 Back 12 Tray Operation 12 Optical Description 13 Computer Connection 13 Q View Software Introduction 13 Chapter 3 Operation 15 Before you Begin 15 Cleaning the Tray Surface...

Страница 6: ...other objects placed in the path of tray travel may be damaged or injured Electricity Since the Q View Imager Pro has a 110V 220V power supply exercise typical precautions to ensure shocks do not occu...

Страница 7: ...to keep any objects such as paper or other flat materials away from the bottom of the Q View Imager Pro as these can block airflow to the fan Space The footprint of the Q View Imager Pro is 26 7 W x 2...

Страница 8: ...and follow the set up wizard prompts 9 Once Q View Software recognizes the Q View Imager Pro green LED lights will turn on around the tray button on the front of the Q View Imager Pro 10 Press the tr...

Страница 9: ...roll Light leak Check Upon completing the Q View Imager Pro installation it may be necessary to ensure there are no light leaks To perform this test remove any plates from the Q View Imager Pro ensure...

Страница 10: ...cords The Q View Imager Pro weighs approximately 13 6 kg 30 lbs Use both hands and lift with your legs not with your back when moving the Q View Imager Pro In order to move or ship the Q View Imager...

Страница 11: ...Biosciences for assistance CHEMICAL SAFETY The only chemicals used with the Q View Imager Pro are deposited within the wells of the plate which is placed on the tray It is important to take precautio...

Страница 12: ...different power cord for any reason please contact technical support Do not use an extension cord for the Q View Imager Pro Ensure that all equipment plugged into the same circuit as the Q View Image...

Страница 13: ...se is 250V 10A T If the device is not powering up when the switch is turned on the user can check the fuse Carefully remove the fuse receptacle Inside the receptacle there are two fuses one is inline...

Страница 14: ...can glow three different colors Green The Q View Imager Pro is on connected to the Q View Software and the tray can be opened Yellow The Q View Imager Pro is acquiring images and the tray is locked R...

Страница 15: ...open if Q View Software is not running the power is turned off or a plate is currently being imaged Once the button is pressed the tray will extend to the point where a plate can be placed on top of...

Страница 16: ...connected to Q View Software when the button glows green From within Q View Software the user can open Imager Management and see an entry in a table for the Q View Imager Pro with one of the columns...

Страница 17: ...ojects Explains Q View Software projects how to create new projects and how to work with multiplex projects at once Part 4 Settings and Preferences Shows how to change your password alter preferences...

Страница 18: ...esidues or smudges are left on the glass or the tray it will adversely affect the quality of the image and the overall quality of your data This can be cleaned with 70 percent ethanol using a lint fre...

Страница 19: ...IMAGING THE PLATE Within Q View Software the user will select Image Processing Then select the tab Acquire Image and under the title Imager select Refresh The name of the Q View Imager Pro will appea...

Страница 20: ...eptable parameters are established by taking the average of the 10 images plus or minus two times the standard deviation of each window read by the Q View Imager Pro To purchase or learn more about th...

Страница 21: ...is overheating Remove airflow obstructions reduce ambient temperature of the room Cannot open tray Q View Imager Pro is not powered on Open Q View Software If that fails ensure drivers are installed...

Страница 22: ...te format with Chemiluminescent blots 7 6 cm x 12 1 cm Imaging Time 1 5 minutes Control Q View Software Connection to PC USB 2 0 Cooling typical Max delta 40 C OPERATING SPECIFICATIONS Operating Condi...

Страница 23: ...20 Q View Imager Pro NOTES...

Страница 24: ...A SPENDLOVE RESEARCH COMPANY 365 North 600 West Logan Utah 84321 T 1 888 782 6797 F 435 750 6869 www quansysbio com Technical Support techsupp quansysbio com 101121GR...

Отзывы: