Quanser QNET Скачать руководство пользователя страница 6

2.1.2 Isolation Amplifier

The HCPL-7800 optical isolation amplifier is used to amplify the electromyogram signal measured by the EMG
electrode, remove noise, and isolate the power source from the user. See the Opto Isolation block in Figure 3.4.
The amplifier has a gain of 8.0 V/V and its output voltage ranges between 1.29 V and 3.8 V. The output of the
isolation amplifier can be measured on A/D #1 when the DIP switch is set to OPTO_OUT.

2.1.3 Muscle Contraction Measurement: EMG Sensor

The EMG Sensor consists of a two-electrode eletromyograph and a grounding strap with a ground electrode. It has
an on-board gain of 300 V/V and a local band-pass filter with lower and upper cutoff frequencies of 25 Hz and 500
Hz, respectively. The electromyogram signal measured by the electromyograph relative to the ground terminal that
is amplified by the isolation amplifier can be measured be measured on D/A #1 by setting the AD1 DIP switch to
OPTO_OUT. The amplitude of the raw EMG signal is small and the signal is offset at around 2.5 V. As shown in
Figure 3.4, the signal is then amplified to fit the

±

10 V range and biased to 0 V. This processed signal is available

on A/D #0 and is used to measure the amount of muscle contraction.

Caution:

Make sure that the two 1.5 V AA batteries that power the EMG sensor are inserted correctly.

2.1.4 DIP Switches

The AD1, AD2, and AD5 DIP Switches dictate what signals can be measured on the Digital-to-Analog lines 1, 2,
and 5, respectively. The AD1 DIP switch, components #6 shown in Figure 3.3, is used to measure the 555 timer or
the output of the optical isolation circuit on D/A #1. As shown in Figure 3.4, the OPTO_OUT is the electromyogram
signal that is measured by the EMG sensor and amplifier and offset by the isolation amplifier. It is offset by about
2.5 V. The 555 Timer resembles a sawtooth wave, but is more like an integrated pulse signal.

Use the AD2 DIP switch, ID #7 in Figure 2.2, to view either DA0_AMP or PRE_STAGE_OFFSET on D/A #2. The
DA0_AMP signal is the processed Digital-to-Analog #0 channel output, as illustrated in Figure 3.4. This is the A/D
#0 signal, i.e. the analog output signal supplied to DAQ, that is scaled down and offset by the post stage offset value
before getting passed to the comparator. The PRE_STAGE_OFFSET is a constant value. It is the offset used to
bring the EMG signal to be around 0 V.

The AD5 DIP switch, component #8 shown in Figure 2.2, determines what signal can be viewed on D/A #5 -
PWM_SIG or POST_STAGE_OFFSET. The PWM_SIG is the pulse-width modulated signal being sent to the servo.
It is the result of passing the 555 Timer pulse and the processed A/D#0 signal through a comparator. The
POST_STAGE_OFFSET is the offset used to regulated the attenuated A/D#0 signal to be about 0 V.

2.1.5 555 Timer

The National Semiconductor LM555CM-ND is a high-precision 555 timer integrated circuit that is used for the PWM
cycles. It can be monitored on A/D#1 by setting AD1 DIP switch to 555_REF.

2.1.6 QNET Power Supply

The DCMCT module has a 15-Volt DC power jack to power the on-board ICs. It is called the QNET Myoelectric
power supply.

Caution:

Please make sure you use the correct type of wall transformer or you will damage the system. It should

supply 15 VDC and be rated at 5.0 A. The QNET Myolectric does

NOT

use the same power supply as other QNET

systems.

Note:

The fuse is replaceable and is rated at 2A 250VAC, type slow-blow, size 5x20 mm.

QNET MYOELECTRIC- User Manual

6

Содержание QNET

Страница 1: ...USER MANUAL QNET Myoelectric Trainer for NI ELVIS Set Up and Configuration Captivate Motivate Graduate...

Страница 2: ...om This document and the software described in it are provided subject to a license agreement Neither the software nor this document may be used or copied except as specified under the terms of that l...

Страница 3: ...4 Specifications 8 5 Environmental 9 6 Setup Guide 10 6 1 QNET and NI ELVIS II Setup 11 7 QNET LabVIEW Hints 14 7 1 Scaling Scopes 14 7 2 Saving Response 15 8 Troubleshooting 18 8 1 General Software...

Страница 4: ...ear envelope filtering zero order hold position control and the basics of LabVIEW coding This is done using a PC with real time control capabilities and the NI ELVIS II The hardware of the MYOELECTRIC...

Страница 5: ...nnector 10 Servo Motor Connector 3 EMG Power On Off SW2 Switch 11 Servo Motor 4 Battery Power Supply for EMG 12 Servo Clamps 5 PCI connector to NI ELVIS for interfacing QNET module with DAC 13 15V QNE...

Страница 6: ...wn in Figure 3 4 the OPTO_OUT is the electromyogram signal that is measured by the EMG sensor and amplifier and offset by the isolation amplifier It is offset by about 2 5 V The 555 Timer resembles a...

Страница 7: ...with a DAQ device is provided in Figure 3 3 The block diagram representing the circuit in the Myoelectric board is shown in Figure 3 4 Figure 3 3 Schematic of QNET Myoelectric trainer Figure 3 4 QNET...

Страница 8: ...5 V Gain 300 V V Upper cut off frequency 500 Hz Lower cut off frequency 25 Hz Common mode rejection ratio 80 dB Supply voltage typical 5 00 V Isolation Amplifier VMAX Recommended input voltage accura...

Страница 9: ...nit Operating temperature 15 to 35 C Humidity 20 to 90 Table 3 MYOELECTRIC environmental operating conditions Caution Ensure the unit is operated under the temperature and humidity conditions given in...

Страница 10: ...setup procedure for using a QNET with an NI ELVIS II Figure 6 5 Connecting a QNET Trainer Caution Do not position the ELVIS II so that it is difficult to disconnect the main power Caution If the equip...

Страница 11: ...he installation procedure are located and marked by an ID number in Figure 6 6 and described in Table 4 6 1 1 ELVIS II Components Figure 6 6 Components on ELVIS II ID Description 1 NI ELVIS II 2 Proto...

Страница 12: ...ELVIS II power cable 4 Connect the ELVIS II USB cable to the PC 5 Connect the supplied QNET transformer to the QNET power jack on the QNET module 6 Connect the myoelectric sensor and grounding straps...

Страница 13: ...6 8 QNET LEDs should all be ON 11 Power ON the myoelectric sensor power 12 Ensure that the Battery ON LED is lit as shown in Figure 6 9 Figure 6 9 Battery LED should be ON QNET MYOELECTRIC User Manual...

Страница 14: ...e steps below to reduce the y axis range of the Angle deg scope shown in Figure 7 10 in order to see the blue trace more up close Figure 7 10 Scope needs to be scaled 1 As illustrated in Figure 7 11 t...

Страница 15: ...0 10 0 However when changing the x axis i e the time scale it is recommended to do the following 1 Pause the scopes or stop the VI and clear the chart right click on scope select Data Operation Clear...

Страница 16: ...e 7 14 opens and gives various image export options One way is to export the image to the clipboard as a bitmap This can then be pasted in a graphical software e g MS Paint Irfanview and saved to a de...

Страница 17: ...5 Figure 7 15 Sample saved response The scope can be saved whether or not the VI is running However typically it is easier to stop the VI when the desired response is collected and then export the ima...

Страница 18: ...wer switch which is located on the back of the ELVIS I and II units and the Prototyping Board Power switch which is situated on the front panel of the ELVIS I and on the top right corner of the ELVIS...

Страница 19: ...ms and select the NI ELVIS II device 4 As shown in Figure 8 16 click on the Reset Device button 5 Once successfully reset click on the Self Test button 6 If the test passed reset the ELVIS II i e shut...

Страница 20: ...pport link Fill in the form with all the requested software and hardware information as well as a description of the problem encountered Also make sure your e mail address and telephone number are inc...

Страница 21: ...QNET 2 0 Rotary Pendulum Board teaches classic pendulum control Quanser QNET add on boards for the NI ELVIS platform teach introductory control topics in undergraduate labs cost effectively All QNETs...

Отзывы: