Table 9. Recommended safety distances between portable and mobile RF
telecommunications devices
Recommended safety distances between portable and mobile RF telecommunications devices
and the QiF
Warrior Extreme
blood and fluid warmer
The QiF Warrior Extreme blood and fluid warmer is intended for use in an electromagnetic
environment in which radiated RF disturbance variables are controlled. The customer or user of the
QiF Warrior Extreme blood and fluid warmer can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the QiF Warrior Extreme blood and fluid warmer as recommended below,
according to the maximum output power of the communications equipment.
Rated
maximum
output power of
transmitter W
Protection distance according to transmitter frequency m
150 kHz to 80 MHz
𝑑𝑑
= 1.17
√𝑃𝑃
80 MHz to 800 MHz
𝑑𝑑
= 0.35
√𝑃𝑃
800 MHz to 2.5 GHz
𝑑𝑑
= 0.7
√𝑃𝑃
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended distance can
be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
9.
Cleaning the System
Cleaning the Compact Disposable Unit
a
The CDU is provided sterile and ready for use.
b
The CDU is a single patient use item; do not reuse it!
c
WARNING
: Do not submerge, sterilize or autoclave the CDU.
Cleaning and disinfecting the Base Unit and Battery
The Warrior EXTREME reusable parts- Base Unit, battery, Extension Cable and
Mount Accessory are supplied non-sterile and should be surface cleaned/disinfected
after each patient use.
Before cleaning, disconnect the BU from the Battery.
Note: the following procedures are not guaranteed to control the spread of pathogens.
Consult the local hospital infection control administrator regarding cleaning
procedure policies at your institution.