TROUBLESHOOTING

CAUTION

 

Opening the OVDX potentially will damage the system and will void any warranty.  If faced with any 
issues with the OVDX, contact your QSA Global, Inc. representative immediately for guidance. 

 

The OVDX has been designed as a rugged, safe, reliable system.  A series of interlocks and self-checks are built into the 
system  to  ensure  safe  operation.    These  checks  may  trigger  an  error  code  which  will  prevent  normal  operation  of  the 
OVDX.  Use Table 4 as a reference guide for system error codes. 

NOTICE:

 Prior to any troubleshooting, please ensure that a fully-charged M18™ battery is installed.  A low battery may 

cause system to report false error codes. 

Error Code 

Description 

Corrective Action 

001 

Memory Failure 

Hard  drive  and/or  system  memory  error  or  failure.    If 
restart does not clear error, contact your QSA Global, Inc. 
representative for repair options. 

002 

LED Failure 

Status  Indicator  LED(s)  are  not  functioning  properly.    If 
restart does not clear error, contact your QSA Global, Inc. 
representative for repair options.  

003 

X-Ray Tube Failure 

X-Ray tube is not communicating or has malfunctioned.  If 
restart does not clear error, contact your QSA Global, Inc. 
representative for repair options. 

004 

X-Ray Over Temperature 

X-Ray  tube  has  exceeded  its  operating  temperature, 
forcing system to stop emitting X-Rays.  Perform following 
actions to recover from this error: 

  Turn off unit 

  Place in cooler environment for 30 minutes 

  Turn unit on 

005 

Board Over Temperature 

Main  control  board  has  overheated.    Perform  following 
actions to recover from this error: 

  Turn Off unit 

  Place in cooler environment for 30 minutes 

  Turn unit on 

006 

Imaging Camera Failure 

Communication error and/or failure with targeting camera.  
If a restart does not clear error, the USB connection may 
have  loosened.    Remove  imaging  housing  from  Carbon 
Fiber  Tube  and  check  USB  connection  to  housing  and 
restart.    If  this  does  not  correct  the  error,  Contact  your 
QSA Global, Inc. representative for repair options.

007 

Targeting Camera Failure 

Communication error and/or failure with targeting camera.  
Contact  your  QSA  Global,  Inc.  representative  for  repair 
options. 

008 

Low Voltage 

M18™ battery has insufficient charge; change battery. 

Содержание OPENVISION DX OVDX-NDT-70

Страница 1: ...SOFTWARE MANUAL MAN 063 REVISION A OPENVISION DX OVDX 70 SOFTWARE MANUAL...

Страница 2: ...ecautions 6 OpenVision Digital X Ray System OVDX Software 7 Power Cycling 7 Startup 7 Shutdown 7 Home Screen 8 Settings Menu 8 Firing X Rays 10 X Ray Live Image Screen 11 Image Capture 12 Video Captur...

Страница 3: ...cm 21 in 58 cm 2 5 in 6 4 cm 3 4 in 8 7 cm Startup Time est 30 seconds Shutdown Time est 5 seconds X Ray Energies Low Medium High Voltage kV 40 55 70 Current mA 0 300 0 218 0 171 Dose 12 in 30 5 cm f...

Страница 4: ...wing standards when used as directed ANSI HPS N43 5 2005 Ionizing Radiation Regulations 2017 ISTA 3A Over the Road vibration standard MIL STD 810 Method 514 Annex C Cat 4 REACH ROHS HMD ANSI Z87 1 US...

Страница 5: ...ng the daily inspection must be removed from service until repaired or replaced The user of this equipment must follow the operating instructions in the order shown to ensure safe operations and compl...

Страница 6: ...ousing imager housing and display s CAUTION Do not attempt to open or modify the imager housing main housing or X ray tube Opening any of these components will void the system warranty and may expose...

Страница 7: ...ts ON position the CPU indicator will immediately turn on green A boot up screen will be displayed during initialization and the System Status Indicators will also illuminate When these indicators tur...

Страница 8: ...y default The Settings menu can be activated from either the Home Screen or the X ray Imaging Screen by holding the hat switch on the controller down towards the round button for three seconds see Fig...

Страница 9: ...settings and user adjustable values available in the Settings menu Setting Units Possible Values X Ray Voltage kV Low Medium High default Low X Ray Current mA Low Medium High default Low Table 2 Setti...

Страница 10: ...o fire X Rays press and hold the oblong palm switch and then the red trigger switch Figure 6 System will transition to the X Ray Warning Screen Figure 5 for five seconds during which the buzzer will p...

Страница 11: ...by default System Status Indicators remain amber colored and the buzzer pulse rate changes to 2 second intervals To change settings while emitting X Rays i e on the fly hold down the hat switch for 3...

Страница 12: ...ome Screen targeting camera image or the X Ray Live Image Screen X Ray image Snapshots are taken by pressing and releasing the round Image Capture button Figure 9 The system will display the snapshot...

Страница 13: ...g the capture button for at least two seconds Within one second the system will commence recording video indicated by a notification Recording in progress appearing in the upper right hand corner of t...

Страница 14: ...Playback Figure 13 Playback Selection Screen The Image Playback on screen button on System Info Screen Figure 12 brings up the Playback Selection Screen Figure 13 Image directories if any are shown o...

Страница 15: ...rly configured USB drive includes an ovdx directory at the drive s top level and an images directory within the ovdx directory see Figure 14 Unless these directories are on the USB drive as indicated...

Страница 16: ...the Media Transfer Screen Figure 15 Navigate to desired folder to copy then press and release the Capture button Figure 3 to initiate file transfer Figure 15 Media Transfer Screen When no image direct...

Страница 17: ...This will transition the user to the Transfer Images confirmation screen Figure 18 NOTICE only one directory of images will be transferred to the USB drive per insertion Following the last step of a...

Страница 18: ...mage directory At the completion of the transfer the system will transition to the delete images screen Figure 20 Image deletion confirmation This screen presents the user with Yes and No buttons used...

Страница 19: ...pace to hold all images this screen will be displayed Selecting the Close button will cause the system to transition to the home screen and the USB drive can be safely removed Software Update Figure 2...

Страница 20: ...he user is presented with a Yes button and a No button Selecting the No button causes the system to transition to the home screen It is safe to remove the USB drive and no software update will be perf...

Страница 21: ...0 kV Current Low 0 090 mA to 0 300 mA Current Medium 0 090 mA to 0 300 mA Current High 0 090 mA to 0 300 mA Current Super 0 090 mA to 0 300 mA Super Mode Enable Disable Require Scatter Shield Enable D...

Страница 22: ...own hat switch will change the value highlighted MAINTENANCE The OVDX was designed to require minimal maintenance for the operator All that is required is a daily visual inspection of the unit Specifi...

Страница 23: ...If restart does not clear error contact your QSA Global Inc representative for repair options 003 X Ray Tube Failure X Ray tube is not communicating or has malfunctioned If restart does not clear err...

Страница 24: ...781 359 9179 qsa global com Error Code Cause Description 009 Camera Communication Error Communication error with internal camera Cycle system power to clear error 010 Camera Communication Error Commun...

Страница 25: ...e LaPorte TX 77571 Telephone 1 713 944 3200 Fax 1 281 476 9309 All goods and services are sold subject to the terms and conditions of QSA Global Inc A copy of these terms and conditions is available u...

Отзывы: