To start recorder for Holter
Step 1
Insert formatted SD Card (using Q200 format) and battery into recorder.
Step 2
Screen will display “Q200/HE” and “QRS Diagnostic” information.* Press ENTER to continue to main
menu. Erase memory if prompted to do so.
Step 3
If desired, adjust Settings by using arrows –
,
and
– to move the cursor and the ENTER button
to select. Use arrows to navigate to General Settings menu and to adjust entries. Press ENTER to
select and return to menu.
Step 4
From main menu go to New Patient screen to verify or enter Patient ID. At least one character must
be entered for Patient ID.
Step 5
Use arrows and ENTER to input ID and press EVENT when finished. (Hold down ENTER button to
backspace.) Recorder will now start automatically after 10 minutes or by pressing EVENT button for
3 seconds. ECG signal and quality will appear on screen.
Step 6
To remove card after recording, gently push inward to release. Never pull card out of slot, as it may
damage the recorder.
*Note:
If screen displays a 15 second countdown, you will need to interrupt by pressing ENTER,
,
, and
EVENT buttons, in that order.
Messages
Lead Loose
– Occurs when the patient is not hooked
up or if there is a problem with the hook up. The
problem may be with an electrode, a lead, or the cable
that connects the leads to the recorder. The message
will continue to flash for about 10 seconds after
corrected. When corrected, the ECG, then Time-of-day
will appear on the screen. If you choose, you can turn
this error message off via the main menu.
Battery LOW
or
FAILURE
– Put in new battery before
starting recorder.
Erase memory YES/NO
– If the SD card has been
used for a previous patient, you will need to erase now.
If the card should not be erased as it has ECG data that
you do not want to lose, you will need to remove
it and put in a new formatted SD card.
SD Card errors
– SD Cards must be formatted using
your LX Holter software. Refer to your Operator’s
Manual for details on SD Card errors and how to format
and correct.
General Settings
Contrast
Adjusts LCD screen contrast
Lead Loose
Enables/Disables “Lead Loose”
error message
Event Marker
Denotes event location on ECG
Key Mode
“Quiet” turns off key beeping.
“Delayed” prevents accidental events
by requiring patient to hold button
down for several seconds to record an
event
Rec Type
Switch between Holter and Event only
recording modes.
Menu Lock
Locks settings on the recorder
To lock recorder enter “217”
To unlock the recorder enter “151”
Language
Choose from multiple languages
Hi Res
Enables/Disables high resolution
recording (Holter only.)
QRS Diagnostic