Rotor-Gene Q MDx CE User Manual 02/2022
159
8
Maintenance
Maintaining the working performance of the Rotor-Gene Q MDx is easy. Optical performance is
maintained by ensuring that the lenses, located at both the emission and detection source, are
clean. This is achieved by gently wiping a cotton tip applicator, moistened with ethanol or
isopropanol*, over the lenses.
Note
: Clean the lenses at least once a month, depending on usage. Wipe the rotor chamber at the
same time.
Keep the work bench area clean and free from dust and sheets of paper. The air inlet of the Rotor-
Gene Q MDx is at the bottom and loose material such as paper or dust may compromise
performance.
To avoid dust build up, keep the lid of the Rotor-Gene Q MDx closed when the instrument is not in
use.
Note
: Only use parts supplied by QIAGEN.
8.1
Cleaning the Rotor-Gene Q MDx surface
The Rotor-Gene Q external surfaces can be cleaned using commonly available laboratory
chemicals.
*
When working with chemicals, always wear a suitable lab coat, disposable gloves, and protective goggles. For more
information, consult the appropriate safety data sheets (SDSs), available from the product supplier.