QIAcuity User Manual 06/2022
24
Optical system
The optical system of the QIAcuity is a camera-based fluorescence microscopy system. The
excitation source for the fluorescence dyes is a high-power white LED. This source in combination
with a specific excitation filter is used to illuminate a whole well at a time. The fluorophores in the
single partitions absorb that light and emit light that is being filtered by a detection filter, collected
and imaged through an objective lens on a CMOS-camera chip (see image below for a detailed
overview of the components). Depending on the configuration of the instrument (2plex/5plex), there
are up to 2 or 5 selectable detection channels. An additional channel is used for detecting the base
fluorescence of the master mix, to determine the exact number of filled partitions and normalization
of fluorescence data.
Scheme of the imaging module.
Available channels
Table 2. Available channels in QIAcuity
Channel
Excitation (nm)
Emission (nm)
Example fluorophores
Green
463–503
518–548
FAM™, EvaGreen
®
Yellow
514–535
550–564
HEX™, VIC
®
Orange
543–565
580–606
TAMRA™
Red
570–596
611–653
ROX™, Texas Red
®
Crimson
590–640
654–692
Cy5
®
Содержание QIAcuity 911000
Страница 2: ...QIAcuity User Manual 06 2022 2...
Страница 68: ...QIAcuity User Manual 06 2022 68...
Страница 126: ...QIAcuity User Manual 06 2022 126 Pressing the Users list button will bring you to the list of all users and their status...
Страница 178: ...QIAcuity User Manual 06 2022 178 Raw image of a well showing dust particles marked with red circles...
Страница 207: ...QIAcuity User Manual 06 2022 207 To open an individual plot click on the appropriate header of the 1D Scatterplot...
Страница 240: ...QIAcuity User Manual 06 2022 240...