
8. Inspection using Resitest-400M
1. Inserting the sponge into the Wenner sensor.
Insert all of the soaked four sponges into the holes of the sensor's
contact. Before inserting the sponges, wash them in tap water or
clean water. During surveying, if the sponge is dried, the sponge
should be soaked in water using bath tub.
2. Connecting the Wenner Sensor to the main body.
After connecting the cable to the Wenner sensor, connect it to
the socket of the main body.
3. Turning on the Resitest-400M.
Power on, using the power key. At that time the Resitest-400M
should be displayed 000 or 001.
4. Confirming the Resitest-400M using the reference block.
When the Wenner sensor of RT-400M is contacted on the reference
confirm if it displays in the range of
62.5
㏀
.cm
±
1.5
or not.
The displayed value in Resitest-400M should be within the range of
61.0 - 64.0
㏀
.
㎝
Resitest 400M-June-2011-mkc-13/28