13
CHAPTER 5
CARING FOR YOUR CAMERA
The Rolera-XR camera requires no regular maintenance except occasional external cleaning of the
CCD window (the glass window between the camera sensor and the microscope or lens).
To clean the CCD window:
CAUTION — The camera’s CCD sensor, and circuits are sensitive to static
discharge. Ensure you are using a static strap or are completely
grounded at all times to release any static energy before you clean the
CCD window.
•
Use clean forced air (available at stores that sell cameras and computer cleaning supplies)
to dust the CCD window.
CAUTION — Do not use acetone.
•
If the image still appears dirty, gently wipe the face of the CCD window with a small
amount of optical grade isopropyl alchohol and lens paper.
•
Apply forced air again to remove any loose particles
.