www.qimaging.com
©2016 QImaging. All rights reserved. QI_Retiga R Series Camera_UM_Rev_A5
16
RETIGA R
™
SERIES USER MANUAL
Camera Fan Not Running
The QI Retiga camera has a fan that operates continuously. If the camera is imaging, but the fan is not
operating, please contact QImaging Support. The fan can be viewed through the vent holes on one side of the
camera.
Camera Running Too Warm
It is normal for the camera to be slightly warm to the touch while in operation. However, if the camera is more
than slightly warm to the touch (and at least one inch of space has been left around the camera housing),
switch off the camera immediately and contact QImaging Support. If the camera has become hot due to
having cooling blocked, move the camera to allow free air flow. Contact QImaging Support if the camera
fails to recover regulated cooling. Note that heat is only emitted from one side of the camera: air goes in the
camera
→
across cooling fins
→
out of camera.