2
Specifications
Table 1.
ST-565 Series Specifications
Specification
*
ST-565
ST-565T
Part Number (P/N)
50664
50671
Azole Species
Halogenated Tolytriazole (HST)
Tolytriazole (TTA)
Azole Range
0–7.5 ppm
0–10.0 ppm
Azole Accuracy
±
0.2 ppm
Calibration
Two-point calibration against standard solution
Outputs
4–20mA Analog Output,
RS-485 Digital Output with Modbus protocol
Installation
Custom tee assembly (P/N: ST-001) with 3/4” FNPT socket & threaded ports
Cable Length
5 ft with IP67 connectors & 2 ft flying lead with IP67 adapter
Power Supply
22–26 VDC, 1 W
Dimension (L
×
Dia)
6.8
×
1.44 inch (172.7
×
36.6 mm)
Weight
0.37 lbs (170 g)
Material
CPVC
Operational
Temperature
32–104 °F (0–40 °C)
Storage Temperature
-4–140 °F (-20–60 °C)
Pressure
Up to 100 psi (0.7 MPa)
Flowrate
0–8 gpm (505 cm
3
/s)
Enclosure Rating
IP67
Regulation
CE
*
With Pyxis’s continuous improvement policy, these specifications are subject to change without notice.
3
Unpacking Instrument
Remove the instrument and accessories from the shipping container and inspect each item for any damage
that may have occurred during shipping. Verify that all accessory items are included. If any item is missing
or damaged, please contact Pyxis Lab Customer Service at
3.1
Standard Accessories
• Tee Assembly 3/4” NPT (1x Tee, O-ring, and Nut)
P/N: ST-001
• 7-Pin Female Adapter/Flying Leads Cable (2 ft)
P/N: MA-1100
• User Manual available online at
https://pyxis-lab.com/support/
ST-565 Series User Manual
| +1 (866) 203-8397
3