ST-565 Series User Manual
[email protected]
| +1 (866) 203-8397
1
Страница 1: ......
Страница 2: ... Series HST or TTA Inline Sensors User Manual October 21 2020 Rev 1 40 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confidential ...
Страница 3: ...onnecting to uPyxis Mobile App 7 5 3 Calibration Screen and Reading 8 5 4 Diagnosis Screen 9 5 5 Device Info Screen 10 6 Setup and Calibration with uPyxis Desktop App 10 6 1 Install uPyxis Desktop App 10 6 2 Connecting to uPyxis Desktop App 11 6 3 Information Screen 11 6 4 Calibration Screen 12 6 5 Diagnosis Screen 12 7 Outputs 13 7 1 4 20mA Output Setup 13 7 2 Communication using Modbus RTU 13 8 ...
Страница 4: ...ivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services for 30 days after the authorization and will correct any qualifying deficiency in labor provided that the labor service deficiency is exactly related to the originating event No other remedy other than th...
Страница 5: ... for by identifying untreated makeup for UV fluorescent signal and removing that from the final value measured The ST 565 Series probe will not work in waters containing Nitrite treatment as Nitrite absorbs UV light and will result in a false high value With these application limitations clearly understood the ST 565 Se ries probe is commonly used for rapid azole validation dramatically reducing s...
Страница 6: ...l Temperature 32 104 F 0 40 C Storage Temperature 4 140 F 20 60 C Pressure Up to 100 psi 0 7 MPa Flowrate 0 8 gpm 505 cm3 s Enclosure Rating IP67 Regulation CE With Pyxis s continuous improvement policy these specifications are subject to change without notice 3 Unpacking Instrument Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have...
Страница 7: ...3 2 Optional Accessories Figure 1 ST 565 Series User Manual service pyxis lab com 1 866 203 8397 4 ...
Страница 8: ...ension of the ST 565 Series probe and the ST 001 Tee Assembly mm 4 2 Wiring If the power ground terminal and the negative 4 20mA terminal in the controller are internally connected non isolated 4 20mA input it is unnecessary to connect the 4 20mA negative wire green to the 4 20mA negative terminal in the controller If a separate DC power supply other than that from the controller is used make sure...
Страница 9: ...e with the uPyxis Mobile App or a computer with the uPyxis Desktop App Figure 3 Bluetooth connection to ST 565 Series probe 4 4 Connecting via USB A USB RS485 adapter P N MA 485 can be used to connect a ST 565 Series probe to a computer with the uPyxis Desktop App NOTE Using non Pyxis USB RS485 adapters may result in permanent damage of the ST 565 Series probe communication hardware Figure 4 USB c...
Страница 10: ...ng to uPyxis Mobile App Turn on Bluetooth on your mobile phone Do not pair the phone Bluetooth to the ST 565 Series probe Open uPyxis Mobile App Once the app is open the app will start to search for the sensor Once the uPyxis Mobile App connects to the sensor press the ST 565 Series probe Figure 6 ST 565 Series User Manual service pyxis lab com 1 866 203 8397 7 ...
Страница 11: ...ult to the Calibration screen From the Calibration screen you can perform calibrations by pressing on Zero Calibration Slope Calibration and 4 20mA Span Follow the screen instructions for each calibration step Figure 7 ST 565 Series User Manual service pyxis lab com 1 866 203 8397 8 ...
Страница 12: ...fines the fluid type that the ST 565 Series probe in currently measuring then press Cleanliness Check If the probe is clean a green Clean message will be shown If the probe is partially fouled a yellow Becoming Dirty message will be shown If the probe is severely fouled a red Dirty message will be shown In this case follow the procedure in the Methods to Cleaning the ST 565 Series Probe section of...
Страница 13: ...upyxis this setup package will download and install the Microsoft Net Framework 4 5 if not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis Setup exe file to install Figure 10 Click Install to start the installation process Follow the screen instructions to complete the US...
Страница 14: ...ormation Screen Once connected to the device a picture of the device will appear on the top left corner of the window and the uPyxis Desktop App will default to the Information screen On the Information screen you can set the information description for Device Name Product Name and Modbus Address then click Apply Settings to save Figure 12 ST 565 Series User Manual service pyxis lab com 1 866 203 ...
Страница 15: ...e reading refresh rate is every 4 seconds Figure 13 6 5 Diagnosis Screen After the device has been calibrated and installation has been completed to check diagnosis click on Diag nosis When in the Diagnosis screen you can view the Diagnosis Condition of the device This feature may be used for technical support when communicating with service pyxis lab com Figure 14 ST 565 Series User Manual servic...
Страница 16: ...configured to provide backup to limit potential product overfeed for example by limiting pump size or duration or by alarming if the pumping rate exceeds a desired maximum limit The ST 565 Series probe is designed to be easily removed inspected and cleaned if required It is suggested that the ST 565 Series probe be checked for fouling and cleaned calibrated on a monthly basis Heavily con taminated...
Страница 17: ...shooting If the ST 565 Series probe output signal is not stable and fluctuates significantly make an additional ground connection connect the clear shield earth ground wire to a conductor that contacts the sample water electrically such as a metal pipe adjacent to the ST 565 Series tee Carry out routine calibration verification against a qualified Halogenated Tolytriazole HST with ST 565 or Tolytr...
Страница 18: ...Contact Us Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com Phone 1 866 203 8397 Email service pyxis lab com ST 565 Series User Manual service pyxis lab com 1 866 203 8397 15 ...