PTSA Measurement
When powered on, the SP-380 will default to the PTSA measurement mode. If the SP-380 is instead in
Fluorescein measurement mode, press
PTSA
(
) to switch to PTSA measurement mode. The water
sample can be transferred to the sample cell using a pipette or filled directly from a faucet, sample bottle,
or sample valve.
*
NOTE
*
Special care should be taken when pouring the sample into the cell to avoid air
bubble entrainment, which can interfere with reading accuracy.
Before beginning a measurement, use the sample water to rinse the sample cell at least three times. Allow
5–10 seconds for the SP-380 to stabilize. The values will be displayed in
white with a blue background
if
a stable value is reached (Figure 2). For a sample containing 100 ppb PTSA, the measured PTSA should be
stabilized within the range of 98–102 ppb.
*
NOTE
*
The time required to reach a stable reading may be slightly longer if the water
sample temperature is significantly different than the environmental temperature at which
the SP-380 had been equilibrated (stored).
6.1.1
High Color and Turbidity Warning
The SP-380 has extra channels to measure sample turbidity and color to automatically compensate sam-
ple color and turbidity interference. If sample turbidity and color values determined are too high, a PTSA
measurement warning will be displayed. In such a case, the user should filter the sample for PTSA mea-
surement.