8.1

Sample Cell Cleanliness Check

The SP-380 is designed to provide reliable and accurate measurement on PTSA and Fluorescein. Heavy
fouling will prevent the light from reaching the sensor, resulting in inaccurate readings. It is suggested
that the SP-380 be checked for fouling and cleaned on a monthly basis. Heavily contaminated waters may
require more frequent cleanings. Cleaner water sources with less contamination may not require cleaning
for several months. The SP-380 is designed to carry out a Cleanliness Check as described below:

1. Power on the SP-380 by pressing

.

2. Press

Info

(

) to launch the

DEVICE INFORMATION

screen.

3. Press

Diagnosis

(

) to launch the

SYSTEM DIAGNOSIS

screen.

4. Allow 5–10 seconds for the message in the top-right corner of the display change from

Starting BTLE...

to

BTLE Started

5. Press

Cleanliness

(

). An instruction prompt appears to ask the user to put DI water into the main

module sample cell (Figure 14).

6. Pour DI water into the main module sample cell.

7. Press

Confirm

(

,

, or

). The instruction prompt will disappear and the SP-380 displays a

countdown toward the bottom of the display.

8. Once the Cleanliness Check is completed a

Clean

message (Figure 15) or

Sample cell fouled

mes-

sage (Figure 16) will appear towards the bottom of the display.

9. Cleanliness check is now complete. Press

Exit

(

) to return to measurement mode.

Figure 14.

Figure 15.

Figure 16.

SP-380 User Manual

[email protected]

| +1 (866) 203-8397

12

Содержание SP-380

Страница 1: ...Images _manualcover jpg...

Страница 2: ...dheld PTSA Fluorescein Dual Meter User Manual December 15 2020 Rev 1 31 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Con...

Страница 3: ...8 Device Information and Diagnosis 11 8 1 Sample Cell Cleanliness Check 12 8 2 Bluetooth Connection to Devices 13 8 3 Factory Reset 15 9 Use with uPyxis Mobile App 16 9 1 Download uPyxis Mobile App 16...

Страница 4: ...elivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services fo...

Страница 5: ...y interference Large color graphic screen that can be read in direct sunlight 2 Specifications Table 1 PTSA Specifications Item Specification Range 0 300 ppb Accuracy 1 or 1 0 ppb Resolution 0 1 ppb E...

Страница 6: ...310 g Humidity 85 at 106 F 41 C Enclosure Rating IP67 With Pyxis s continuous improvement policy these specifications are subject to change without notice Batteries excluded 3 Unpacking Instrument Re...

Страница 7: ...dheld Cleaning Kit SER 02 4 Installation 4 1 Battery Installation The SP 380 is powered by four alkaline batteries Typical battery life lasts for 10 000 measurements or 10 months When the battery capa...

Страница 8: ...he SP 380 has three control keys as shown in Figure 2 The left right and ok keys are used to launch actions indicated on the LCD display directly above the keys The labels above the keys indicate the...

Страница 9: ...The values will be displayed in white with a blue background if a stable value is reached Figure 2 For a sample containing 100 ppb PTSA the measured PTSA should be stabilized within the range of 98 10...

Страница 10: ...nning a measurement use the sample water to rinse the sample cell at least three times Allow 5 10 seconds for the SP 380 to stabilize The values will be displayed in white with a blue background if a...

Страница 11: ...art the zero blank calibration 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 5 7 Rinse the sample cell three times with the desired PTSA stand...

Страница 12: ...he DI water 2 Power on the SP 380 by pressing Allow 5 10 seconds for the SP 380 to stabilize 3 Press FC to switch to Fluorescein measurement mode 4 Press to launch the FLCN CALIBRATION screen Figure 7...

Страница 13: ...nt calibration or b Press Exit to end the calibration process at a three point calibration and return to Fluo rescein measurement mode 7 2 3 Four Point Calibration 15 Rinse the main module sample cell...

Страница 14: ...or main module also shown Press Diagnosis to launch the SYSTEM DIAGNOSIS screen where raw measurement data are displayed Figure 13 The information has no use for normal operation but instead is used f...

Страница 15: ...INFORMATION screen 3 Press Diagnosis to launch the SYSTEM DIAGNOSIS screen 4 Allow5 10secondsforthe messagein the top rightcornerof thedisplaychangefrom Starting BTLE to BTLE Started 5 Press Cleanline...

Страница 16: ...ter taken from the inline sensor sample line The SP 380 can then be used to calibrate the inline sensors over the Bluetooth connection To calibrate an inline sensor follow the steps below 1 Power on t...

Страница 17: ...ly the same as the value measured by the SP 380P ress Read again to take more readings from the sensor if nec essary 14 If the calibration is successful a Verify calibration OK message will appear on...

Страница 18: ...to BTLE Started 5 Press Help to launch the HELP screen Figure 23 6 Press Factory Reset or The display updates as shown in Figure 24 and the user can choose one of three options a Press to start the f...

Страница 19: ...le smart phone according to the following steps 1 Follow the steps in the Bluetooth Connection to Devices section to make the SP 380 discoverable 2 Open uPyxis Mobile App 3 On uPyxis Mobile App pull d...

Страница 20: ...hen connected the uPyxis Mobile App will default to the Setting screen From the Setting screen the user can set the Power off time and Screen off time in seconds Figure 27 SP 380 User Manual service p...

Страница 21: ...e press Get Firmware Once the new firmware is downloaded press Upgrade Firmware NOTE The firmware update process takes some time and will require the SP 380 to stay within range approximately 10 ft wi...

Страница 22: ...not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis Setup exe file to i...

Страница 23: ...ices section to make the SP 380 discoverable 2 Plug the Bluetooth USB adapter into a USB port in the computer 3 Launch uPyxis Desktop App 4 On uPyxis Desktop App click Device Connect via USB Bluetooth...

Страница 24: ...ate firmware file contact service pyxis lab com for these firmware files and clicking Upgrade Firmware NOTE The firmware update process takes some time and will require the SP 380 to stay within range...

Страница 25: ...month The SP 380 will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Read Datalog deleted by clicking Delete or exported by clicking Expo...

Страница 26: ...dark will reset the timer and the screen will re light This press will not step along the calibration process the next press needed will still be required in the sequence Always rinse the unit with cl...

Страница 27: ...re condition such as leaving the SP 380 inside an unattended automobile 12 Contact Us Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com Phone 1 866 203 8397 Email service...

Отзывы: