6.5
Diagnosis Screen
After the device has been calibrated and installation has been completed, to check diagnosis, click on
Diag-
nosis
. When in the
Diagnosis
screen you can view the Diagnosis Condition of the device. This feature may
be used for technical support when communicating with
. To preform a Cleanliness
and Selflife Check, first select the
Diagnosis Condition
which defines the fluid type that the ST-772/T sensor
is currently measuring, then click
Cleanliness and Selflife Check
. If the sensor is clean, a
Clean
message
will be shown. If the sensor is fouled, a
Please replace the DO membrane cap
message will be shown. In
this case, follow the procedure in the
Methods to Cleaning the ST-772/T
section of this manual.
Figure 15.
7
Outputs
7.1
4–20mA Output Setup
The 4–20mA output of the ST-772 sensor is scaled as:
• Dissolved Oxygen:
–
4 mA = 0 ppm
–
20 mA = 20 ppm
7.2
Communication using Modbus RTU
The ST-772/T sensor is configured as a Modbus slave device. In addition to the ppm dissolved oxygen value,
many operational parameters, including warning and error messages, are available via a Modbus RTU con-
nection. Contact Pyxis Lab Customer Service (
) for more information.
ST-772/T User Manual
| +1 (866) 203-8397
17