Troubleshooting
If the ST-525SS probe output signal is not stable and fluctuates significantly, make an additional ground
connection – connect the clear (shield, earth ground) wire to a conductor that contacts the sample water
electrically such as a metal pipe adjacent to the ST-525SS tee.
Carry out routine calibration verification against a qualified Fluorescein standard. After properly cleaning
the ST-525SS sensor, carry out the zero point calibration with distilled water and slope calibration using the
qualified Fluorescein standard.