ST-525SS Inline Fluorescein Fluorometer

Probe

User Manual

March 20, 2020

Rev. 3.00

Pyxis Lab, Inc.

1729 Majestic Dr. Suite 5

Lafayette, CO 80026 USA

www.pyxis-lab.com

© 2017 Pyxis Lab, Inc.

Pyxis Lab Proprietary and Confidential

Содержание 50666

Страница 1: ...Inline Fluorescein Fluorometer Probe User Manual March 20 2020 Rev 3 00 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Con...

Страница 2: ...pp 6 5 2 Connecting to uPyxis Mobile App 6 5 3 Calibration Screen and Reading 7 5 4 Diagnosis Screen 7 5 5 Device Info Screen 8 6 Setup and Calibration with uPyxis Desktop App 8 6 1 Install uPyxis Des...

Страница 3: ...livery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services for...

Страница 4: ...ile or Desktop App Any standard containing Fluorescein in the range of 10 to 60 ppb can be used for the calibration of the ST 525SST he standard may also be the water sample itself if the Fluorescein...

Страница 5: ...Leads Cable 2 ft P N MA 1100 User Manual available online at www pyxis lab com support html 3 2 Optional Accessories Figure 1 4 Installation The ST 525SS probe has two 3 4 female NPT ports It is reco...

Страница 6: ...the controller is used make sure that the output from the power supply is rated for 22 26 VDC 65mA NOTE The negative 24V power terminal power ground and the negative 4 20mA ter minal on the ST 525SS p...

Страница 7: ...adapters may result in permanent damage of the ST 525SS probe communication hardware Figure 4 USB connection to ST 525SS probe 5 Setup and Calibration with uPyxis Mobile App The ST 525SS probe can be...

Страница 8: ...le App Turn on Bluetooth on your mobile phone Do not pair the phone Bluetooth to the ST 525SS probe Open uPyxis Mobile App Once the app is open the app will start to search for the sensor Once the uPy...

Страница 9: ...may be used for technical support when communicating with service pyxis lab com To preform a probe cleaniness check first select the Diagnosis Condition which defines the fluid type that the ST 525SS...

Страница 10: ...25SS probe can be calibrated in a two point zero slope procedure using a deionized water sam ple and a standard containing 10 to 60 ppb Fluorescein The calibration solution could be the sample water i...

Страница 11: ...package from http www pyxis lab com sup port html this setup package will download and install the Microsoft Net Framework 4 5 if not previously installed on the PC the USB driver for the USB Bluetoo...

Страница 12: ...information description for Device Name and Product Name then click Set to save Figure 12 6 4 Calibration Screen To calibrate the device click on Calibration On the Calibration screen there are three...

Страница 13: ...ion on how to calibrate then click Slope Calibration Figure 14 6 4 3 4 20mA Span To perform 4 20mA Span click on 4 20mA Span Then follow the instructions provided to alter the 4 20mA output span of th...

Страница 14: ...Lab Customer Service service pyxis lab com for more information 8 Sensor Maintenance and Precaution The ST 525SS probe is designed to provide reliable and continuous Fluorescein readings even when in...

Страница 15: ...ning Solution Kit P N SER 01 which can be purchased at our online Estore Catalog https pyxis lab com product st series probe cleaning kit Figure 17 Inline Probe Cleaning Solution Kit To clean the ST s...

Страница 16: ...the ST 525SS tee Carry out routine calibration verification against a qualified Fluorescein standard After properly cleaning the ST 525SS sensor carry out the zero point calibration with distilled wat...