4.2
ST-500SS Piping
The ST-500SS sensor has 3/4” female NPT threaded ports on the sensor itself and therefore does not require
a custom tee assembly. It is recommended that two 3/4” NPT to 1/4” tubing adapters are used to connect
the sensor to the sampling system. Sample water entering the sensor must be cooled down to below 104
°F (40 °C). The sensor can be held by a 1.75-inch pipe clamp or mounted to a panel with four 1/4-28 bolts.
See Figure 3 for ST-500SS dimensions.
Figure 4.
Dimension of the ST-500SS (inch)