In-line Piping
The provided ST-001 Tee Assembly can be connected to a pipe system through the 3/4” female ports, ei-
ther socket or NPT threaded. The ST-001 Tee Assembly should be placed in-line with the suction side of the
chemical feed pump. This method should ideally in a vertical line run to avoid chlorine gas bubble interfer-
ence. Pyxis also offers 2” and 3” inline tee assemblies as an accessory option if required. To properly install
the ST-600 Series sensor into the ST-001 Tee Assembly, follow the steps below:
1. Insert the provided O-ring into the O-ring groove on the tee.
2. Insert the ST-600 Series sensor into the tee.
3. Tighten the tee nut onto the tee to form a water-tight, compression seal.
Figure 2.
Dimension of the ST-600 Series sensor and the ST-001 Tee Assembly (mm)
ST-600 Series User Manual
| +1 (866) 203-8397