To clean the ST-600 Series sensor, soak the lower half of the sensor in 100 mL inline sensor cleaning solution
for 10 minutes. Rinse the ST-600 Series sensor with distilled water and then check for the flashing blue light
inside the ST-600 Series sensor quartz tube. If the surface is not entirely clean, continue to soak the ST-600
Series sensor for an additional 10 minutes. Use the small, soft bristle brush and Q-Tips cotton swabs as
necessary to remove any remaining contaminants in the ST-600 Series sensor quartz tube.
8.2
Storage
Avoid long term storage at temperature over 100 °F. In an outdoor installation, properly shield the ST-600
Series sensor from direct sunlight and precipitation.
9
Troubleshooting
If the ST-600 Series sensor output signal is not stable and fluctuates significantly, make an additional ground
connection –– connect the clear (shield, earth ground) wire to a conductor that contacts the sample water
electrically such as a metal pipe adjacent to the ST-600 Series tee.
Carry out routine calibration verification against a qualified bleach standard. After properly cleaning the
ST-600 Series sensor, carry out the zero-point calibration with distilled water and slope calibration using the
qualified bleach standard.