pyroscience FireStingO2-Mini-ST Скачать руководство пользователя страница 17

14 

each time when pin 7 of connector X1 is tied to ground (triggered 
auto-mode).  The  results  of  the  measurements  can  be  either  read 
from  the  transmit  pin  of  the  UART  interface,  or  from  the  analog 
outputs at connector X2. 

NOTE:  When  the  auto-mode  is  activated,  the  receive  pin  of  the 
UART  interface  has  no  function.  The  module  will  not  respond  to 
any command sent to it. 

The basic idea behind the auto-mode is that a fully configured and 
calibrated  module  can  be  operated  without  the  need  of  complex 
communication  protocols.  However,  the  configuration  (including 
the configuration of the auto-mode!) and the calibration data have 
to  be  adjusted  in  the  full-control  mode  (or  by  using  the  USB-
interface  cable  in  combination  with  the  comfortable  windows 
software 

Pyro  Oxygen  Logger

;  please  refer  to  the  next  chapter). 

The communication protocol offers a special command for storing 
the  current  configuration and  calibration  data permanently  in  the 
internal  flash  memory.  Thus,  a  fully  configured  module  can  be 
operated in auto-mode even if the power supply is intermediately 
interrupted. 

A  typical  application  scenario  would  be  the  following:  the 

FireStingO2-Mini-ST 

could  be  configured  and  calibrated  in  a 

"calibration  lab"  using  the  full-control  mode  (or  the  windows 
software 

Pyro

 

Oxygen  Logger

).  The  configuration  and  the 

calibration data are stored in the internal flash memory. Then the 
sensor is brought to the "field application" which can be a remote 
place.  By  connecting  pin 6  of  the  connector  X1  to  ground  the 
module  starts  measuring  without  the  need  of  any  UART 
communication.  The  measurement  results  can  be  read  out  from 
the analog output (or from a text-string sent via the UART transmit 
line). 

Содержание FireStingO2-Mini-ST

Страница 1: ...FireStingO2 Mini ST OPTICAL OXYGEN OEM MODULE WITH UART INTERFACE ANALOG OUTPUT FOR FIBER OPTIC SENSORS MANUAL...

Страница 2: ...ireStingO2 Mini ST is manufactured by PyroScience GmbH Hubertusstr 35 52064 Aachen Germany Phone 49 0 241 5183 2210 Fax 49 0 241 5183 2299 Email info pyroscience com Internet www pyroscience com Regis...

Страница 3: ...EXTERNAL TEMPERATURE SENSOR 7 4 ELECTRICAL CONNECTORS 9 4 1 CONNECTOR X1 POWER DIGITAL INTERFACE 10 4 2 CONNECTOR X2 ANALOG OUTPUT 11 5 OPERATION VIA THE UART INTERFACE 12 5 1 FULL CONTROL MODE 12 5...

Страница 4: ...e FSO2 Mini ST is compatible to most oxygen sensors types available from PyroScience Further it includes also internal high precision sensors for i ambient atmospheric pressure ii relative humidity of...

Страница 5: ...ts inside the device The FireStingO2 Mini ST and the sensors should be kept and stored outside the reach of children in a secure place under dry and clean conditions at room temperature avoiding moist...

Страница 6: ...ibration and application of the sensors is on the user s authority as well as data acquisition treatment and publication The sensors and the oxygen meter FireStingO2 Mini ST are not intended for medic...

Страница 7: ...indicated on the type label Just penetrate the type label at this position with e g a needle or nail in order to access the M3 thread 3 1 Optical Port for Oxygen Sensors The FSO2 MINI ST provides an...

Страница 8: ...5 Fig 1 Drawing of the FSO2 MINI ST...

Страница 9: ...t air However the readings of the internal temperature sensor should be interpreted more carefully First the internal heating by the electronics typically increases the internal temperature by about 1...

Страница 10: ...ilized for such a compensation the first choice should be always an external 4 wire PT10o temperature sensor placed directly within the sample where the oxygen concentration is measured The FireStingO...

Страница 11: ...s do not get into contact to the aluminum housing NOTE In order to minimize potential electrical noise coupling into the external temperature sensor the 4 wire cable should be twisted and as short as...

Страница 12: ...without any soldering or crimping When inserting or removing a stripped cable end stripping length 6 mm max core diameter 0 5 mm into one of the connector holes of the connectors S1 or S2 an internal...

Страница 13: ...VCC Power Power supply min 3 5V max 5 0V DC max 50 mA typical 30 mA 3 GND Power Ground 4 TXD Digital Output 0V or 3 3V Data transmission pin of the UART interface 5 RXD Digital Input 0V or 3 3V 5V to...

Страница 14: ...n analog output is configured as an alarm output it will give 2 5V if a specific parameter is out of a defined range and 0V if it is within this range For further details please refer to the separatel...

Страница 15: ...for further information However after power cycling the module always reverts to the standard baud rate of 19200 There are two principal operation modes possible 1 the full control mode and the 2 auto...

Страница 16: ...control mode each single measurement has to be triggered by a specific command sent via the UART interface e g the MSR or TMP commands for measuring oxygen or temperature respectively Alternatively an...

Страница 17: ...software Pyro Oxygen Logger please refer to the next chapter The communication protocol offers a special command for storing the current configuration and calibration data permanently in the internal...

Страница 18: ...lly the USB interface cable FSO2 MINI USB provides also a virtual COM port Custom Windows software can use this virtual COM port for communicating directly with the module based on the communication p...

Страница 19: ...older version please download and install the newest version Please refer to the manual provided with the software for general operation instructions In combination with OEM applications it is importa...

Страница 20: ...est way would be to operate the module in the auto mode see previous chapter But even for using it in an OEM application under the full control mode can make sense because the communication protocol f...

Страница 21: ...measuring principle lifetime detection of REDFLASH indicator luminescence Oxygen channel connector fiber optic ST plug of FireStingO2 Mini ST Excitation wavelength 620 nm orange red Detection waveleng...

Страница 22: ...1 C Accuracy 0 3 C Range 40 to 125 C Response Time ca 1 minute Internal Humidity Sensor Resolution 0 04 rel humidity Accuracy typ 0 2 rel humidity Range 0 to 100 rel humidity Response Time t63 ca 10 s...

Отзывы: