
P/N: M-145MAT00-20
Rev. A -
Page Rev.A
III-20
3) Flow sensor calibration
3-a) Set up material
1-Remove the top housing (refer to the disassembly section page III-22),
2-Locate on the MCU board the trimmers TR3 and TR4,
3-Realise the test set up figure 3.1 page III-4
4-Connect the unit to the computer,
5-Follow the steps described page III-5 to III-6 and click on “Calibrations/Performances”,
the following menu will appear.
TR4
TR3
GoodKnight 425 / 425 ST
Service Manual
Содержание GoodKnight 425
Страница 1: ...GoodKnight 425 SERVICE MANUAL GoodKnight 425ST P N M 145MAT00 20 Revision A...
Страница 2: ......
Страница 8: ...GoodKnight 425 425ST Service Manual P N M 145MAT00 20 Rev A Page Rev A 6...
Страница 9: ...I GENERAL INFORMATION GoodKnight 425 425ST Service Manual P N M 145MAT00 20 Rev A Page Rev A I 1...
Страница 15: ...II THEORY OF OPERATION P N M 145MAT00 20 Rev A Page Rev A II 1 GoodKnight 425 425ST Service Manual...
Страница 23: ...III SERVICE AND REPAIR GoodKnight 425 425 ST Service Manual P N M 145MAT00 20 Rev A Page Rev A III 1...
Страница 39: ...P N M 145MAT00 20 Rev A Page Rev A III 17 Personal Notes GoodKnight 425 425 ST Service Manual...
Страница 50: ...P N M 145MAT00 20 Rev A Page Rev A III 28 Personal Notes GoodKnight 425 425 ST Service Manual...
Страница 51: ...IV SPARE PARTS GoodKnight 425 425ST Service Manual P N M 145MAT00 20 Rev A Page Rev A IV 1...
Страница 55: ...GoodKnight 425 425ST Service Manual P N M 145MAT00 20 Rev A Page Rev A...