UVC LAMP REMOVAL OR REPLACEMENT
The Shield 1500 has been designed such that the UVC Lamp can be serviced without
removing the sheet metal cover of the Shield 1500.
1.
Disconnect power to Shield 1500
and disconnect
both the water input and output connections.
2. Remove easy-off end cap by pulling cap off static
gland nut as shown in the top image of Figure 15.
3. Carefully withdraw lamp approximately 2 inches
from chamber. While holding lamp end, remove lamp
socket from exposed lamp end as shown in the
middle image of Figure 15.
4. Carefully withdraw lamp from chamber. Be sure to
withdraw lamp straight out without angling until
completely clear of quartz sleeve as shown in the
bottom image of Figure 15.
CAUTION: Lamp and quartz sleeve are easily
damaged. Exercise care when handling.
5. Reinstall lamp in reverse order.
WARNING: Germicidal ultraviolet rays are harmful to eyes and skin. Do not
restore power to the Shield 1500 until the UVC lamp and easy-off end caps have
been properly reinstalled.
Figure 15. Removing UV lamp