52
Immunity
Test
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines.
IEC61000-4-11
<5%UT
(>95% dip in
UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle, 6
cycle
70% UT
(30% dip in UT)
for 25 cycle, 30
cycle
<5% UT
(<95% dip in
UT)
for 5s
<5%UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle, 6
cycle
70% UT
(30% dip in UT)
for 25 cycle, 30
cycle
<5% UT
(<95% dip in UT)
for 5s
Main power quality should
be that of a typical
commercial or domestic
environment.
If the user of the Milk Genie
requires continued
operation during power
interruptions, it is
recommended to power
the pump from an
uninterruptible power
supply or the battery.
Power Frequency
(50/60
㎐
)
Magnetic Field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
commercial or domestic
environment.