Battery Function
The device is equipped with a battery to ensure the maintenance of operations for at least 30 minutes without
mains power supply.
The PiCCO
2
must be connected to the mains power supply for at least 6 hours in order to completely charge
the battery.
In order to maintain the charging condition of the battery, the device must remain on mains power supply
during storage and the main switch at the back of
the device must remain on position “ON”.
The battery symbol in the information bar shows the current state of charge:
Battery charging
Grey, last bar of status display is blinking
Fully charged, device connected to
mains power
No battery symbol
Battery-powered, > 25% capacity
Grey, remaining capacity in 25% steps
Battery-powered, < 25% capacity
Red, blinking with acoustic alarm
No battery in device
Grey, crossed out symbol
In cases of a battery alarm the acoustic alarm can be paused for 5 minutes.
NOTE
If the operating time of the fully charged battery drops below 30 minutes, or the remaining
operating time after the battery alarm has started (a red blinking battery symbol) drops below 7
minutes, contact your PULSION representative to replace the internal battery.
Ensure that the batteries are disposed of correctly after expiration of battery life.
NOTE
In case of a totally discharged battery the message “Battery totally discharged” appears on the
screen. To ensure a fully charged battery, the device must be connected to mains power supply
for up to 24 hours.
CAUTION:
Check the battery state of charge when using the device. Only when the battery is fully
charged, the use of the device can be ensured for the stated time without being connected to the