WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
Ensure that the service tasks described below
are only performed by specialized staff.
In addition, ensure that all other service tasks
are only performed by PTW-Freiburg or by a
person authorized by PTW-Freiburg.
7 Cleaning
General Information on Cleaning
NOTICE
To avoid surface damage, do not use disinfec-
tants on a phenol base or peroxide compounds
to disinfect the surface.
The product cannot be sterilized or completely
disinfected.
The product must not be treated or come in con-
tact with fat, oil, alcohol, or substances containing
solvents.
Cleaning agents and solvents contained in disin-
fectants (alcohols) have proved to be particularly
harmful.