Follow these instructions to clean the emitter points:
1. Remove the front grill.
2. Moisten a swab or wiping cloth in the IPA solution.
3. Swab or wipe the emitter points until they are free of
particles.
4. Make certain the emitter points are straight and parallel
to the plane of the fan.
5. Replace the front grill.
6. Make sure the emitter points are dry before powering the
ionizer again.
Emitter Point Replacement
The IN5110 uses 10 tungsten alloy precision etched
emitter needles. Contact Transforming Technologies for in-
formation about ordering replacement emitters. The part
number for emitters is listed below:
Replacement part
Part Number
Tungsten replacement emitters
22-6423
Replacement Filter
22-6450
Because the IN5110 high voltage output is AC, emitter
erosion from the ionization process on the electrodes is
minimal. Unless physically broken or stressed, the IN5110
emitters should last the life of the ionizer.
Follow these instructions to remove the emitter points:
1. Turn off and disconnect the unit from the AC power.
2. Remove the front grill.
3. Unscrew the threaded emitter point using a needle nose
pliers.
4. Replace and tighten the new emitter using the same
tool. Do not over-tighten.
5. Repeat steps 3 and 4 as required.