PT W80158 Скачать руководство пользователя страница 2

 FEATURES

  WARNINGS & MAINTENANCE

• Refractometers is a field device for the measurement of an index of refraction.  
• Used to calculate the composition of various automobile fluids.
• Measures DEF, Ethylene Glycol, Propylene Glycol, Battery Fluid, and Cleaning Fluids.
• Precision instrument features high-quality optics with ultra-smooth focusing
• ATC - Automatic Temperature Compensation
• Heavy duty lightweight aluminum construction, includes a protective storage case.

1.  Accurate measurement depends on careful calibration. The prism and 
 

sample must be at the same temperature for accurate results.

2.  Do not expose the instrument to damp working conditions, and do not
 

immerse the instrument in water. If the instrument becomes foggy, water has

 

entered the body. Call a qualified service technician or contact you dealer. 

3.  Do not measure abrasive or corrosive chemicals with this instrument. They 
 

can damage the prism’s coating.

4.  Clean the instrument between each measurement using a soft, damp cloth. 
 

Failure to clean the prism on a regular basis will lead to inaccurate results 

 

and damage to the prism’s coating.

5.  This is an optical instrument; it requires careful handling and storage. 
 

Failure to do so can result in damage t the optical components and its basic 

 

structure. With care, this instrument will provide years of reliable service. 

 OPERATION

  PARTS DIAGRAM

Main Prism 

Assembly

Calibration 

Screw

Focus 

Adjustment

Eyepiece

Rubber Grip

Daylight

Plate

Step 1

Open daylight plate and place 2 – 3 drops of distilled water on the main prism. Close 
the daylight plate so the water spreads across the entire surface of the prism without air 
bubbles or dry spots. Allow the sample to temperature adjust on the prism for approxi-
mately 30 seconds before going to step # 2. (This allows the sample to adjust to the 
ambient temperature of the refractometer). 

Step 2

Hold daylight plate in the direction of a light source and look into the eyepiece. You will 
see a circular field with graduations down the center (you may have to focus the eyepiece 
to clearly see the graduations). The upper portion of the field should be blue, while the 
lower portion should be white. (The pictures shown here and shown in step 3 & step 4 are 
only as reference the right specific scale is listed the product.) 

POOR POOR

GOOD

1.070

1.060

1.050

1.040

1.030

1.020

1.010

1.000

100

90

80

70

60

50

40

30

20

10

0

d

20

20

%

0

1.070

1.060

1.050

1.040

1.030

1.020

1.010

1.000

100

90

80

70

60

50

40

30

20

10

0

d

20

20

%

0

Calibrate to “0”

Reading of Sample

As seen when looking 

into the instrument

Step 3

Look into the eyepiece and turn the Calibration Screw until the 
boundary between the upper blue field and the lower white 
field meet exactly on the zero scale, such as shown in the 
image. That is the end of the calibration process. Make sure the 
ambient room temperature is correct for the solution you are 
using (20° C/ 68° F). When working temperature of the room 
or environment (not the sample) changes by more than 5 F, we 
recommend recalibrating to maintain accuracy. If the instrument 
is equipped with Automatic Temperature Compensation 
system, the ambient working temperature of the room must 
be 20° C (68° F) whenever the instrument is recalibrated. Once 
calibrated, shifts in ambient temperature within the acceptable 
range (10° C- 30° F) should not affect accuracy. 

Step 4

Now place a few drops of the ample to be tested onto the main 
prism, close the daylight plate and check reading. Take 
the reading where the boundary line of blue and white cross the 
graduated scale. The scale will provide a direct reading 
of the concentration. 

Light 

Source

Отзывы: