Minox i Oxygen Transmitter User Manual
14
ProcessSensing.com
Appendix B - Hazardous area certification
Area
Certification details
Standards
Europe & Japan
(Pending)
IECEx / ATEX / UKCA
IECEx BAS 19.0013 / Baseefa 19ATEX0020 /
BAS22UKEX0052
II 1 GD
Ex ia IIC T4 Ga (-20 °C≤Ta≤+55 °C)
Ex ia IIC
T200
135°C Da (-20 °C≤Ta≤+55 °C)
Ui=28 V Ii=93 mA Pi=0.65 W Ci=12nF Li=705uH
E
N IEC 60079-0:2018
EN 60079-11:2012
IEC 60079-0 Ed. 7 2017
IEC 60079-11 Ed. 6 2011
North America /
Canada
(Pending)
Class I Division 1 Group A-D T4
Class II Division 1 Groups E-G T135°C
Class I Zone 0 AEx ia IIC T4 Ga / Ex ia IIC T4 Ga
Class 1 Zone 20 AEx ia IIC T135°C Da / Ex ia IIC
T135°C Da
Ta = -20°C≤Ta≤+55°C
UI/Vmax=28 V II/Imax=93 mA Pi=0.65 W Ci=12nF
Li=705uH
CSA C22.2 No. 60079-0:19
CSA C22.2 No. 60079-11:14
UL 60079-0:2019
UL 60079-11:2014
CSA C22.2 No. 61010-1-12