Product User Guide | 3
MicroRHTemp
PRODUCT USER GUIDE
Device Operation (cont’d)
Alarm Settings
To change the settings for the alarm:
1. In the
Connected Devices
panel, select the intended
device to change the alarm settings.
2. On the
Device
tab, in the
Information
group, click
Properties
. Users can also right-click on the device and
select
Properties
in the context menu.
3. Click
Alarm
. A window will appear allowing to set the
high and low alarms and warning alarms.
4. Check
Enable Alarm Settings
to enable the feature
and check each high, low and alarm box to activate it.
The values can be entered in the field manually or by
using the scroll bars.
5. Click
Apply
to apply the changes to the device. Click
Save
to save the changes to a template.
6. To clear an active alarm, press the
Clear Alarm
button.
7. To set an alarm delay, enter the duration of time into
the
Alarm Delay
box in which the readings can be
outside of the alarm parameters.
LEDs
Once started, the green LED will flash at the selected
reading rate to indicate that the device is running. The red
LED will flash in one second intervals if there is an alarm
condition.
Device Maintenance
Battery Replacement
Materials:
SR1154W Battery (2 required)
1. Unscrew the knurled end cap.
2. Tip the batteries (enclosed in a plastic sleeve) out of
the enclosure tube.
3. Use a small, dull, non-metallic tool (e.g. pen cap) to
push the batteries out of the sleeve.
4. Press the new batteries into the sleeve negative (-) end
first.
5. Place the sleeved batteries in the enclosure tube
positive (+) end first.
6. Screw the knurled end cap back in place.
Recalibration
MadgeTech recommends annual recalibration. To send
devices back for calibration, visit
www.pacificsensortech.com.au